• Digital Interface: USB 2.0
• Outputs: Scalable 4-20 mA, 0-20 mA, 0-10 V, 0-5 V, Type J, K, R or S Thermocouple.
• Inputs: Digital inputs for emissivity control, ambient background temperature
compensation, trigger/hold input.
• Alarm Relay: 48 VAC, 300 mA, optically isolated.
• Accuracy: ±4°F (2°C) for target temp. <68°F (20°C), else larger of ±1% of reading
or ±2°F (1°C) for mA output, or ±1% of reading or ±5°F (2.5°C) for Thermocouple.
• Repeatability: ±0.5% of reading or ±1°F (0.5°C), whichever is greater.
• Temperature Resolution (mA output): ±0.2°F (±0.1°C).
• Cable Length: Compatible with 1m (3.3ft) standard, 3m (10ft), 8m (26ft), 15m (50ft),
and 30m (100ft) lengths.
• Output Impedance (T/C output): 20 ohms.
• Minimum Load Impedance (mV output): 10K ohms.
• Maximum Load Impedance (mA output): 500 ohms.
• Power Supply: 8 to 32 VDC, max. 4W.
• Housing: NEMA 4 (IP65) Rated Zinc Die-Cast Electronics Box.
• Operational Ambient Temperature: 14 to 150°F (-10 to 65°C).
• Storage Temperature: -4 to 185°F (-20 to 85°C).
• Relative Humidity: 10 to 95%, non-condensing.
• EMI/EMC/ESD: IEC EN61326-1 1:2006.
• Shock: IEC 68-2-27: 50 G, 11 ms, 3 axes.
• Vibration: IEC 68-2-6: 3 G, 11-200 Hz, 3 axes.
• Weight: 13 oz (370 g).