Typical Applications |
Features |
- Mass spectrometry
- Electron beams
- Test equipment
|
- High precision and high stability
- 15PPM temperature coefficient
- 0 to 5VDC
- 0 to 3 Amps of current
- Maximum Iout capability down to 0 Volts
- Programmable voltage and current controls
- Indefinite output short-circuit protection
- Buffered output current & voltage monitors
- Excellent linearity & accuracy of control
- Current mode and voltage mode indicator
- Synchronizable
|
Input
Parameter |
Conditions |
All types/models |
Operating range (VDC) |
All conditions |
+24 ± 10 |
Current (mA) |
Full load output |
900mA typical |
Output
Voltage range (VDC) |
Nominal input |
0 to 5 |
DC current range (amps) |
Nominal input |
0 to 3 |
Voltage range (VDC) |
Derated |
0 to 5.7 |
DC current range (amps) |
Derated |
0 to 3.3 |
Voltage monitor scaling (VDC) |
Full load |
10 |
Current monitor scaling (VDC) |
Full load |
10 |
Programming and Controls
Input impedance (MΩ) |
Nominal input |
+ Output models 10MΩ to GND |
Adjust resistance (Ω) |
Typical potentiometer values |
10K to 100K (Pot across Vref. & Signal GND, Wiper to Adjust) |
Adjust voltage (VDC) |
Referenced to signal ground |
0 to +10 VDC |
Accuracy (amps) |
In current control |
±0.1% |
Offset (VDC) |
Voltage control |
0.04% |
Offset (amps) |
Current control |
0.001% |
Output voltage (VDC) |
T=+25°C, Initial Value |
+10.0V ± 0.05% |
Enable/disable (VDC) |
|
0 to +0.5 Disable, +2.4 to 10 Enable (Default = Enable) |
Environmental
Operating (°C) |
Full load, max Eout, case temp. |
+10 to +45 |
Coefficient (PPM/°C) |
Over the specified temperature |
|
Thermal shock (°C) |
Mil-Std 810, method 503-4, Proc. II |
-40 to +65 |
Storage (°C) |
Non-operating, case temp. |
-55 to +85 |
Altitude |
Standard package, all conditions |
Sea level through vaccuum |
Shock (G's) |
Mil-Std-810, Method 516.5, Proc. IV |
20 |
Vibration (G's) |
Mil-Std-810, Method 514.5, Fig.514.5C-3 |
10 |
|