- The IGUU 2916 covers the complete frequency range from 9 kHz - 1 GHz. Obviously both relay contact generators with their 1-dB amplitude setting (main generators) are included with slight modification, because there is still no alternative.
- New Auxiliary Generators open the pulse frequency range beyond 200 Hz. So finally this limit of the mechanical relay contact has been overcome. Pulse frequency in the range of 0.1 Hz to 20 kHz is now available. This means complete testing of emi-receivers including detectors for cispr-quasi-peak, average and peak/mil.
- Rotary encoders and led-displays are used for pulse frequency and amplitude (main generator only) tuning and reading. Pulse frequency generation uses a crystal reference generator and dividers.
- The Sine Wave Generator provides the frequencies 100 kHz, 1 MHz, 10 MHz and 100 MHz with 60 dBmV. The generator uses output voltage control and is referenced to the crystal generator using frequency dividers and multipliers. This means sine wave/pulse comparison without an external signal generator.
- The optional Coaxial Switching Unit KU 9616 connects one of the four outputs of the IGUU 2916 to the e. u. t., controlled by the IGUU 2916. This is very convenient and time saving because there is no need to change coaxial connections.
- The optional IEEE-Interface serves for complete PC-control of all generator functions. Automatic testing is possible by combining interface and coaxial switching unit.
- Pulse frequency range: 0.1 Hz-200 Hz 0-20 Hz in 0,1 Hz-Steps, 0 -200 Hz in 1 Hz-steps
- Pulse frequency accuracy: better +/ -2x10 -4
- Pulse shape: Almost rectangular
- Pulse duration: Band A: 270 ns, Band B: 6 ns, Band C/D: 0,3ns
- Pulse polarity: positive/negative (switchable)
- Pulse level: (CISPR weighting Q.P.): Band A/B 10 -70dB, in 1 dB-steps, Band C/D 0 -60 dB, in 1 dB-steps
- Max. Output Voltage acr. 50 W : 80 V - 110 V
- Pulse Area at 60 dB attenuator setting: Corresponds to CISPR 16 Standard equivalent to 1mVrms
|