With high sample rates of 50 MS/s (GX1200) and 100 MS/s (GX1201), the GX1200 Series is an ideal modulation source for troubleshooting encoding schemes. Both models also provide high speed waveforms to stimulate signal distortion, power line cycle dropouts, video signals, components failures and power supply transients.

The GX1200 Series comes standard with 2 M samples of waveform memory, accessible via a high speed interface. Waveform memory is segmentable. This allows for storage and access of up to 4,096 different waveforms rapidly without having to download waveforms multiple times, which in turn enhances test throughput.


Thông số kỹ thuật


STANDARD WAVEFORMS
Sine, triangle, square, pulse, ramp, sinc (sine(x)/x), Guassian pulse, exponential fall, rising pulse, noise, DC
Frequency Range Waveform dependent
Source Internal synthesizer
SINE
Frequency Range 100 µHz to 25 MHz (GX1200)
100 µHz to 50 MHz (GX1201)
Band Flatness 5% to 10 MHz; 20% to 50 MHz (GX1200)
5% to 10 MHz; 20% to 25 MHz (GX1201)
Programmable Parameters Start phase, 0 °C to +360 °C
Harmonics and Non-Related Spurious at 5 VPP< -55 dBc for carrier frequencies 1 MHz
< -45 dBc for carrier frequencies 5 MHz
< -35 dBc for carrier frequencies 10 MHz
< -28 dBc for carrier frequencies 25 MHz (GX1200)
< -22 dBc for carrier frequencies 50 MHz (GX1201)
THD 0.05% to 100 kHz