FEATURES
Designed for high throughput surface characterization
21 CFR Part 11 Compliant software (optional).
Six independent analysis stations.
Each analysis station served by a dedicated Po station, which increases the accuracy of pore size results.
Bench Top or Cart Mounted Capability: Flexibility in space.
Improved communications & data capabilities.
64 Bit software compatibility - Fully
Compatible with Windows® 7+ and Vista.
Windows software with superior data reduction capabilities.
PERFORMANCE FEATURES:
Improved designs for operator safety.
Built in filters at each station minimize the potential for elutriating powders into the analysis system.
Unlike some other analyzers, a compressed air supply is not required - just mains electricity and analysis grade gases, plus coolant.
Each sample port accepts three standard cell diameters and a variety of shapes to accommodate a wide range of sample types.
The Dewars handles make them easily removable, yet they sit securely on the instrument because of their platforms’ raised edges and unique alignment feature.
The smooth Dewar elevator system assures years of trouble-free operation
Expert users can program each requested datum point independently for equilibration.
Automatically generated log files detail each analysis for easy troubleshooting by Quantachrome’s experienced service staff.
A meaningful instrument status “dashboard” indicates analysis progress point-by-point for each station and latest updated Po value for each Dewar.
Intuitive Windows-based software prompts the user for essential information that might be missing before an analysis can be started.
The data reduction program can handle data files from other Quantachrome instruments for easy sharing of results.
Tabular results can be saved as text files and exported as csv files for use in popular spreadsheet programs.
Software allows single-point BET to be calculated (and plotted) at more than one P/Po value at one time. Allows for “what-if” approaches, i.e. what would the single point BET value be if it was calculated at another P/Po value.
Report generation and printing can be fully automated, and customized page-by-page for content and order. A user has the flexibility to do BET before isotherm, or the other way round.
PRESSURE (Specifications by
transducer manufacturer ) |
Range: 0 - 1,000 torr
|
|
Accuracy: |
0.1% full scale |
|
Linearity: |
0.5% full scale |
|
Pressure Resolution: |
0.00026 torr |
|
Minimum P/Po: (N2) |
0.0001 |
ULTIMATE VACUUM:
|
5 x10-3 mbar ( two stage rotary pump - included). |
|
ADSORBATES:
|
Nitrogen or any non-corrosive gas with appropriate coolant including Ar, CO2 and H2 |
SURFACE AREA RANGES:
|
0.01 m2/g to no known upper limit |
PORE SIZE RANGE:
|
3.5 to >4000Å / 0.35 to 400nm
|
MINIMUM PORE VOLUME
(Liquid):
(STP):
|
1 x10-6cc/g
5 x10-5cc/g
|
|
ACCESSORIES: |
|
Required(not included): |
Adsorbate and helium (minimum purity 99.999%), coolant (eg. LN2), PC (Windows), degasser (XeriPrep Degasser, MasterPrep, FloVac or Flow Degasser). |
|
DIMENSIONS: |
|
Width |
29.5 inches (64.8 centimeters) |
Height |
44.5 inches (101.6 centimeters) |
Depth |
37.1 inches (73.7 centimeters) |
WEIGHT: |
380 pounds (172.7 kilograms) |
ELECTRICAL: |
100-240 volts, 50/60 Hertz |
|
AMBIENT: |
10 - 38°C. operating range |
|
90% maximum relative humidity
|
|
|
|
|