Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Thiết bị đo điện trở tĩnh điện PAS-853BRM hãng Prostat

Mã sản phẩm: PAS-853BRM
Sử dụng cho Model:
Hãng SX: Prostat
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

The PAS-853BRM Digital Surface Resistance Test Kit is the ideal kit for the ESD Plant Auditor

Liên Hệ Hỗ Trợ

SĐT: 0984 843 683

    • Mail: info@ttech.vn

    • Zalo: 0984.843.683

Thông tin sản phẩm

CONTROLS
ON/OFF Battery Bus Cut-off
Slide Switch ON SIDE OF CASE


ON: Energizes instrument and LCD displays Software Version then GO A with AUTO indication (Default)


OFF: Isolates battery and DC voltage from instrument controls and circuit. Use for transport and storage


TEST Pad
With Test Leads installed: Press once & Release.
Measurements performed automatically


In conjunction with MODE changes from Auto operation to Continuous


MODE Pad
Steps through Auto/Voltage ranges with each press of MODE Pad:


AUTO: Default automatic resistance range & Voltage control
<10V: Manual Operation 0.01 to =10V Test Voltage;
Resistance Range: 0.01E-1 to 9.99E9 O
10V: Manual Operation at 10V (Nominal, Semi-Constant)
Resistance Range: 0.98E4 to 9.99E10 O
100V: Manual Operation at 100V (±<5V) Constant Voltage
Resistance Range: 0.98E6 to 9.99E12 O
When data in HOLD, press MODE to terminate HOLD without changing operational mode
Press MODE with TEST to change from Auto to Continuous
Used to measure lead resistance (NULL)
Used by Prostat Authorized Technicians for Voltage Calibration
CONNECTIONS
Positive (+): Test Voltage (RED) – Supplies test voltage to electrodes/fixtures


Negative (-): Current sensing (BLACK) – Conducts current from electrode, fixtures or Ground to instrument


Chassis: Shielding connection to chassis (WHITE) – Reduces transmitted energy and effect on current measurement


Note: ¾ inch spacing between RED and BLACK accommodates dual banana BNC adapters


INDICATION
Colored LED’s indicate instrument operational MODE and Test Voltage (YELLOW) used for the current measurement, as well as measurement completion (GREEN LED).


AUTO: ON during automatic operations; OFF in Manual mode
<10V
10V
100V
GREEN LED: Measurement hold
TYPICAL TOLERANCES
1Nominal performance by range:


0.01E- 1 – 1.0E 0 O <±10%, ±2 Counts
1.0E 0 – 9.99E10 O <±2%, ±2 Counts
1.0E11 – 9.99E12 O <± 10%, ±4 Counts