RANGE
Optimal Resistance Range from 0.1 (1.0E-1) Ohms to 1 Tera ohms (1.0E+12 ohms).
TEST VOLTAGES
Automatic Mode (Default): 0.01 volts to 10 volts Variable 1.0E-1 to 1.0E+4 Ohms
Constant Voltage:
10 volts: ± <0.2 volts 1.0E+4 to <1.0E+6 Ohms
100 Volts: ± <2.0 volts 1.0E+6 to >1.0E+12 Ohms
Manual Mode: 0.01 volts to 10 volts Variable 1.0E-1 to <1.0E+5 Ohms
10 volts: ± <0.2 volts 1.0E+2 to 1.0E+9 Ohms
100 volts: ± <2.0 volts 2.0E+5 to >1.0E+12 Ohms
ACCURACY
Overall: ± <5% at ambient conditions (at 23°C and 30% Rh).
Specific Range Tolerances:
1.0E-1 to 1.0E+1 ohms: ±5% Corrected for Test Lead Resistance
1.0E+1 to 1.0E+12 ohms: ±2.0% with 10-foot Test Leads
1.0E+12 to 2.0E+14 ohms: <±40%
DISPLAY
Multi-function 2-5/8” x 1-5/8” Liquid Crystal Display with 0.5" digit height.
Displays 3-1/2 digits in Ohms, or 1.0EXX in exponential format.
Ohms Display indicators: Ω, KΩ, MΩ, GΩ and TΩ. Includes 19-segment analog scale (1-10 with 0.5 indication) with x1, x10, & x100 multipliers.
Number of Data Points in Memory (0 – 80).
Automatic Electrification Time (seconds), or Time required to Manually obtain steady state measurement.
Displays data HOLD, BATTERY status, MIN, MAX, AVG, REC and Test Voltage (<10, 10, or 100 V)
LED INDICATORS
14 Color LED’s from <103 to >1012 ohms.
Colors (RED, GREEN, YELLOW/ORANGE, or Blank/OFF)
TIMER
Times measurements in Seconds up to 99 seconds (Displayed on LCD)
MEMORY
Register stores up to 80 data points (MEM # Displayed after RESET)
|