                                 RANGE 
Optimal Resistance Range from 0.1 (1.0E-1) Ohms to 1 Tera ohms (1.0E+12 ohms). 
TEST VOLTAGES 
Automatic Mode (Default): 0.01 volts to 10 volts Variable 1.0E-1 to 1.0E+4 Ohms 
Constant Voltage: 
10 volts: ± <0.2 volts 1.0E+4 to <1.0E+6 Ohms 
100 Volts: ± <2.0 volts 1.0E+6 to >1.0E+12 Ohms 
 
 
Manual Mode: 0.01 volts to 10 volts Variable 1.0E-1 to <1.0E+5 Ohms 
10 volts: ± <0.2 volts 1.0E+2 to 1.0E+9 Ohms 
100 volts: ± <2.0 volts 2.0E+5 to >1.0E+12 Ohms 
 
 
ACCURACY 
Overall: ± <5% at ambient conditions (at 23°C and 30% Rh). 
Specific Range Tolerances: 
1.0E-1 to 1.0E+1 ohms: ±5% Corrected for Test Lead Resistance 
1.0E+1 to 1.0E+12 ohms: ±2.0% with 10-foot Test Leads 
1.0E+12 to 2.0E+14 ohms: <±40% 
 
 
DISPLAY 
Multi-function 2-5/8” x 1-5/8” Liquid Crystal Display with 0.5" digit height.  
Displays 3-1/2 digits in Ohms, or 1.0EXX in exponential format.  
Ohms Display indicators: Ω, KΩ, MΩ, GΩ and TΩ. Includes 19-segment analog scale (1-10 with 0.5 indication) with x1, x10, & x100 multipliers.  
Number of Data Points in Memory (0 – 80).  
Automatic Electrification Time (seconds), or Time required to Manually obtain steady state measurement.  
Displays data HOLD, BATTERY status, MIN, MAX, AVG, REC and Test Voltage (<10, 10, or 100 V) 
 
 
LED INDICATORS 
14 Color LED’s from <103 to >1012 ohms. 
Colors (RED, GREEN, YELLOW/ORANGE, or Blank/OFF) 
TIMER 
Times measurements in Seconds up to 99 seconds (Displayed on LCD) 
MEMORY 
Register stores up to 80 data points (MEM # Displayed after RESET) 
                             |