Specification:
Industry : Process
Gas : Oxygen
Application : Industrial Gas, Glove Boxes
Measuring cell type : Electrochemical: Percentage, trace and RACE™ cell (US & UK patents)
Measurement range : Trace 0.1ppm - 1%. RACE™ 0.1ppm - 30%. Percent 0.3% - 100%
Response Time : Trace 90% within 30 sec. RACE™ Air to 20ppm within 2 min. Percent 90% within 30 sec.
Unsuitable gases : Acid gases, corrosives and solvents in significant concentration
Features
Long life, maintenance-free disposable oxygen sensor
Ambient air or traceable gas calibration
Microprocessor controlled functions
Large, autoranging LED display
Fast response. Ultra fast response version
Unaffected by vibration or position
Sturdy, reliable construction with three sensor options
Specific to oxygen measurement
Insensitive to sample flow rate - percentage through ppm
Nitrosave flushing gas control option EC9500
The EC900 has a 36 month warranty which covers any faulty workmanship and normal component failure relating to electronic circuit cards
Applications
Chemical/Petrochemical - Chemical production, High purity gas production, Hydrocarbon refining, Natural gas transmission
Curing - Electron beam, Ultraviolet
Electronics - Reflow/wave soldering, Solder powder production, Semiconductor furnaces, Gas quality
Metals - Heat treating/Annealing, Steel production, Alloys and powdered metals
Pharmaceutical - Inert packaging, Vessel blanketing, Fermentation
Process - Ceramics, Combustion analysis, Contact lens manufacturing, Food packaging, Glass fibre optics, Inert gas welding, Lamp manufacturing
General - Controlled environments, R&D, Glove boxes, Oxygen deficiency