| General |
| Enclosure |
19” Rack System, H= 2.1m |
| Power Supply |
100 to 115 V or 220 to 240 V 50/60Hz |
| Pressure Swing Dryer |
| Gas Output |
Flow: 7 Nl/min
Pressure: 0.68 barg
Moisture Content: <1ppmV (<-75°Cdp) |
| Required Gas Supply |
Flow: 10 Nl/min
Pressure: 5 to 7 barg
Moisture Content: Oil and liquid water free |
| Type |
Twin column desiccant, pressure swing |
| Desiccant |
4 Ångström Molecular sieve bead (4-8 mesh) |
| Timer |
Motorized cam |
| Operating Temperature |
+5 to +35°C |
| Generator |
| Generation Range |
Humidity: 1 to 95% RH (-50 to +50°Cdp)
Temperature: Dependant on temperature chamber |
| Generated Gas Output |
Air 2 Nl/min @ 0.5 barg via heat traced line |
| Dual Stage MFC Mixing |
Dual mass flow controllers |
| Power Consumption |
550 V A maximum |
| Dual Stage MFC Mixing |
Dual mass flow controllers |
| Operating Temperature |
+5 to +40°C; 10 to 90% RH |
| Enclosure |
19" Rack System, H= 2.1m |
| Control System |
Closed loop feedback |
| Reference Instrument |
| Performance |
| Measurement Technology |
Chilled Mirror |
| Measurement Range |
-80 to +85°Cdp |
| Measurement Accuracy |
±0.1°Cdp
±0.1° temperature |
| Repeatability |
Better than ±0.1°C |
| Resolution |
0.01 (0.1 for % RH) |
| Dew-Point Sensor |
| Mirror |
Gold plated copper |
| Temperature Measurement |
4 wire Pt100, 1/10 DIN class B |
| Sample Flow Rate |
0.1 to 0.7 Nl/min (recommended) |
| Integrated Flowmeter |
0 to 1 Nl/min |
| Sensor Pressure |
Atmospheric |
| Configuration |
Remote |
| Remote PRT |
| Temperature Measurement |
4 wire Pt100, 1/10 DIN class B |
| Monitor |
| Resolution |
0.01°C |
| Dual Optics Detection |
Wide band red LED with dual photo sensors, all system insulated |
| Outputs |
Analog: 2 channels, 10 mV/°Cdp, 4-20 mA
Digital: RS232
Logic: Data hold, ABC Logic status, optics alarm |
| Operting Temperature |
0 to + 40°C |