MEASURING |
Range: |
Accuracy: |
Resolution: |
Milliamps mA |
0...24 mA (Impedance ±17 Ohm) |
0.02% FS |
1 µA |
Millivolts mV |
-10...+100 mV (> 1 MegaOhm) |
0.01% FS |
1 µV |
Thermocouples |
Types J, K, T, E, N, U, L |
0.1% FS *) |
0.1°C |
Thermocouples |
Types B, R, S |
1% FS *) |
0.1°C |
Resistance
****) |
0...400.00 Ohm
0...2200.0 Ohm |
0.05% FS
0,05% FS |
0.01 Ohm
0.1 Ohm |
RTD
****) |
Types Pt50, Pt100, Pt200, Pt500,
Pt1000, Ni100, Ni120 |
0.1% FS |
0.01°C |
Continuity |
with a 100 Ohm trigger. Visual and audible confirmation |
SOURCING |
Range: |
Accuracy: |
Resolution: |
Milliamps mA |
0...24 mA (max. load 500 Ohm) |
0.02% FS |
1 µA |
Millivolts mV |
-10...+100 mV (min. load 100 Ohm) |
0.01% FS |
1 µV |
Thermocouples |
Types J, K, T, E, N, U, L |
0.1% FS *) |
0.1°C |
Thermocouples |
Types B, R, S |
1% FS *) |
0.1°C |
Resistance |
10.00...400.00 Ohm **)
400.0...2200.0 Ohm **) |
0.05% FS **)
0.05% FS **) |
0.01 Ohm
0.1 Ohm |
RTD |
Types Pt50, Pt100, Pt200, Pt500,
Pt1000, Ni100, Ni120 |
0.1% FS |
0.01°C |
General Data |
Isolation: |
Max. 100 VDC between all input and output circuits |
Thermocouple Extender: |
Cold junction compensation Pt100 DIN EN 60751, 100 Ohm at 0°C,
nom. resistance drift 0.04% after 1000 h at 500°C |
Power supply |
Type: Lithium-Ion battery
Charging time: approx. 5 h
Life time: typical 8 h (depending on used functions) |
Mains charger: |
Charger input at calibrator: 10...15 VDC
Charger input: 115...230 VAC; 0.5 A; 50/60 Hz
Charger output: 12 VDC at 1.5 A
|
Optional car charger: |
12 VDC at 1 A via cigarette lighter plug |
Temperature ranges: |
Storage: -40...+50°C incl. battery
Usage: 0...50°C |
Relative humidity: |
0...85% r.F. (non-condensing) |
Housing: |
ABS-plastic OL94V-0 flame retardant, IP 54 dust and splash proof |
Rubber boot: |
green |
Dimension: |
without rubber boot 149 x 34 x 77 mm
with rubber boot 155 x 43 x 86 mm |
Weight: |
approx. 340 g |