Model |
CS9917AX |
CS9917BX |
CS9917CX |
CS9917DX |
Test Mode |
AC |
DC |
AC/DC |
AC/DC |
ACW |
Output voltage
|
Range |
0.500kV~10.00kV |
|
0.500kV~10.00kV |
Accuracy |
±(2%reading+5V) |
|
±(2%reading+5V) |
Resolution |
1V |
|
1V |
Max output power |
500VA |
|
200VA |
500VA |
Test current |
50mA |
|
20mA |
50mA |
Low-current |
0~49.9mA |
|
0~19.99mA |
0~49.9mA |
Range |
200uA、2mA、20mA、50mA |
|
200uA、2mA、20mA |
200uA、2mA、20mA、50mA |
Waveform |
Sine wave |
Distortion |
≤2%(Unloaded or resistance of the load) |
Crest factor |
1.3~1.5 |
Waveform mode |
DDS+Amplifier |
Rise time |
0,0.3s~999.9s 0=OFF |
Test time |
0,0.3s~999.9s 0=Continuous test |
Fall time |
0,0.3s~999.9s 0=OFF |
Interval time |
0.0s~999.9s 0=OFF |
Output mode |
N、G |
DCW |
Output voltage |
Range |
|
0.500kV~10.00kV |
Accuracy |
|
±(2%reading+5V) |
Resolution |
|
1V |
Max. output power |
|
200W |
100W |
Max. Test current |
|
20mA |
10mA |
Range |
|
2uA、20uA、200uA、2mA、10mA(20mA) |
Creast factor |
|
≤5%(6kV/range) |
Discharge time |
|
≤200ms |
Max. charge current |
|
20mA |
10mA |
Rise time |
|
0,0.3s~999.9s 0=OFF |
Test time |
|
0,0.3s~999.9s 0=Continous test |
Fall time |
|
0,0.3s~999.9s 0=OFF |
Interval time |
|
0.0s~999.9s 0=OFF |
Delay alarm |
|
0,0.3s~999.9s 0=OFF |
Output mode |
|
N、G |
Volt Meter |
Range |
0.500kV~10.00kV |
Accuracy |
±(2%reading+5V) |
Resolution |
1V |
Displayed value |
RMS |
Current Meter |
Range |
AC |
0 ~ 50mA |
DC |
0 ~ 20mA |
Resolution |
AC |
200uA:0.1uA,
2mA:1uA,
20mA:10uA
50mA:100uA |
DC |
2uA: 0.001uA,
20uA: 0.01uA,
200uA:0.1uA,
2mA: 1uA,
20mA: 10uA |
Accuracy |
≥2mA:±(2%+5counts),<2mA:±(3%+5counts) |
Offset |
Current of test leads and accessories can be deducted. |
Mode |
GND:Return port is connected to the case.
FLOAT: RETURN port is not connected to case
|
Timer |
Range |
0~999.9s |
Resolution |
0.1s |
Accuracy |
±(0.1%+50ms)
|