Thiết bị phân tích khí ZKJ Fuji Electric

Mã sản phẩm: ZKJ
Sử dụng cho Model:
Hãng SX: Fuji Electric
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

Simultaneous and continuous measurement of up to 5 components among: NO, SO2, CO, CO2, N2O, CH4, and O2


Ideal for flue gas monitoring in incinerators and industrial furnaces

The mass flow type twin detector — hardly affected by interference gas
A variety of functions: auto calibration, O2-corrected output, averaging, alarm, etc.
Easy-to-see LCD
You can check the concentration changes of all components at a glance
19-inch rack mount
Built-in paramagnetic O2 sensor or optional zirconia O2 sensor
Maximum range ratio 1 : 25
A user can set or change a range within the limit

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

Specifications

Principle NO, SO2, CO, CO2, N2O, CH4: Non-dispersive infrared sensor
O2: built-in paramagnetic sensor or external zirconia sensor (ZFK7, optional) 
Components 1 component: NO, SO2, CO, CO2, CH4, N2O
2 components: NO + SO2, NO + CO, CO2 + CO, N2O + CO2
3 components: NO + SO2 + CO, N2O + CO2 + CO
4 components: NO + SO2 + CO + CO2, NO + SO2 + N2O + CO2
CH4 can be combined.
O2 measurement can be added to above all. 
Range NO: 0–50 ppm…5000 ppm
SO2: 0–50 ppm…10%
CO: 0–50 ppm…100%
CO2: 0–20 ppm…100%
N2O: 0–200 ppm…2000 ppm
CH4: 0–200 ppm…100%
O2: 0–5%…25%
Repeatability ±0.5% FS (±1% FS for 50 ppm or less)
Zero drift ≥ 200 ppm range: ≤ ±1.0% FS per week
50–200 ppm range: ±2.0% FS per week
≤ 50 ppm range: ±2.0% FS per day
Span drift ≤ ±2.0% FS per week
(50 ppm or less: ±2.0% FS per day)
Linearity ±1.0% FS
Response time ≤ 60 s (for 90% response) at a sample gas flow rate of 0.5 L/min
Output signal • 4–20 mA DC (linear) or 0–1 V DC (linear)
(instantaneous values, O2-corrected instantaneous values, O2-corrected average values, O2 average) 
• Contact output 
(CO peak count alarm, concentration alarm, under auto calibration, device error, range identification, calibration error, etc.) 
Contact input Auto-calibration start, average reset, range switch, average reset, pump ON/OFF
Functions Auto calibration, O2 correction, averaging, CO peak count and alarm, output signal hold
Operating environment -5 to +40°C, RH 90% or less
Power supply voltage 100–240 V AC, 50/60 Hz, 250 VA
Dimensions 483 (H) × 177 (W) × 578 (D) mm
Gas conditions Temperature 0–50°C
Flow rate 0.5 L/min ±0.2 L/min
Dust ≤ 100 μg/Nm3 with particle size ≤ 1 μm
Pressure ≤ 10 kPa (gas outlet should be open to the atmospheric air)
Mist Unallowable
Corrosive gas ≤ 1 ppm
Moisture Below the saturation vapor density at 2°C (Non condensing)