Principle |
NO, SO2, CO, CO2, N2O, CH4: Non-dispersive infrared sensor
O2: built-in paramagnetic sensor or external zirconia sensor (ZFK7, optional)
|
Components |
1 component: NO, SO2, CO, CO2, CH4, N2O
2 components: NO + SO2, NO + CO, CO2 + CO, N2O + CO2
3 components: NO + SO2 + CO, N2O + CO2 + CO
4 components: NO + SO2 + CO + CO2, NO + SO2 + N2O + CO2
CH4 can be combined.
O2 measurement can be added to above all.
|
Range |
NO: 0–50 ppm…5000 ppm
SO2: 0–50 ppm…10%
CO: 0–50 ppm…100%
CO2: 0–20 ppm…100%
N2O: 0–200 ppm…2000 ppm
CH4: 0–200 ppm…100%
O2: 0–5%…25% |
Repeatability |
±0.5% FS (±1% FS for 50 ppm or less) |
Zero drift |
≥ 200 ppm range: ≤ ±1.0% FS per week
50–200 ppm range: ±2.0% FS per week
≤ 50 ppm range: ±2.0% FS per day |
Span drift |
≤ ±2.0% FS per week
(50 ppm or less: ±2.0% FS per day) |
Linearity |
±1.0% FS |
Response time |
≤ 60 s (for 90% response) at a sample gas flow rate of 0.5 L/min |
Output signal |
• 4–20 mA DC (linear) or 0–1 V DC (linear)
(instantaneous values, O2-corrected instantaneous values, O2-corrected average values, O2 average)
• Contact output
(CO peak count alarm, concentration alarm, under auto calibration, device error, range identification, calibration error, etc.)
|
Contact input |
Auto-calibration start, average reset, range switch, average reset, pump ON/OFF |
Functions |
Auto calibration, O2 correction, averaging, CO peak count and alarm, output signal hold |
Operating environment |
-5 to +40°C, RH 90% or less |
Power supply voltage |
100–240 V AC, 50/60 Hz, 250 VA |
Dimensions |
483 (H) × 177 (W) × 578 (D) mm |
Gas conditions |
Temperature |
0–50°C |
Flow rate |
0.5 L/min ±0.2 L/min |
Dust |
≤ 100 μg/Nm3 with particle size ≤ 1 μm |
Pressure |
≤ 10 kPa (gas outlet should be open to the atmospheric air) |
Mist |
Unallowable |
Corrosive gas |
≤ 1 ppm |
Moisture |
Below the saturation vapor density at 2°C (Non condensing) |