Features
· Output ON operation is accepted only when the output voltage is 0V. Also, the output voltage is set
to 0 V automatically by output OFF operation.
· Since the operation of the tester is carried out by the contact signal, it is possible to operate even from a place away from the equipment and the test specimen.
· Since the test time can be set arbitrarily, output can be turned off automatically at the end of the test.
· Suppresses heat generation by using insulated oiled transformer and heat radiation of product is done by natural air cooling, so
there is no noise due to mounting of FAN motor
| Input voltage |
1 φ 100 V ± 10% |
| Output voltage |
1 φ 0 ~ 25 kV rms error: 25 kV ± 1.5% |
| Voltage variable system |
Motor drive system |
| Cutoff current |
5-10
- 20 mArms (switching) error: 20 mA ± 2.5%
Response time: about 100 msec |
| capacity |
500VA |
| frequency |
Depends on input |
| Rating |
Half an hour |
| Buck-boost speed |
0.92 kV / sec at input 1φ 100 V |
| timer |
Multitype effective range: 2 sec - 30 min
30 min or more is not guaranteed. |
| input signal |
Output ON, Output OFF, Boost, Buck, Timer |
| Output signal |
During output, test OK, test NG, voltage monitor 1,
voltage monitor 2, HIGH LIMIT, LOW LIMIT |
| Appearance dimension |
W 500 x H 1196 x D 677 |
| mass |
Approximately 180 kg |
| Other specifications |
Operation unit integrated type With
zero - cross switch With
limit switch (HIGH, LOW) |
|