Features
· Output ON operation is accepted only when the output voltage is 0V. Also, the output voltage is set
to 0 V automatically by output OFF operation.
· Since the operation of the tester is carried out by the contact signal, it is possible to operate even from a place away from the equipment and the test specimen.
· Since the test time can be set arbitrarily, output can be turned off automatically at the end of the test.
· Suppresses heat generation by using insulated oiled transformer and heat radiation of product is done by natural air cooling, so
there is no noise due to mounting of FAN motor
Input voltage |
1 φ 100 V ± 10% |
Output voltage |
1 φ 0 ~ 25 kV rms error: 25 kV ± 1.5% |
Voltage variable system |
Motor drive system |
Cutoff current |
5-10
- 20 mArms (switching) error: 20 mA ± 2.5%
Response time: about 100 msec |
capacity |
500VA |
frequency |
Depends on input |
Rating |
Half an hour |
Buck-boost speed |
0.92 kV / sec at input 1φ 100 V |
timer |
Multitype effective range: 2 sec - 30 min
30 min or more is not guaranteed. |
input signal |
Output ON, Output OFF, Boost, Buck, Timer |
Output signal |
During output, test OK, test NG, voltage monitor 1,
voltage monitor 2, HIGH LIMIT, LOW LIMIT |
Appearance dimension |
W 500 x H 1196 x D 677 |
mass |
Approximately 180 kg |
Other specifications |
Operation unit integrated type With
zero - cross switch With
limit switch (HIGH, LOW) |
|