Withstand voltage test part test voltage |
Output voltage |
AC 0.2kV - 5.0kV |
Voltage output method |
Inverter type |
Transformer capacity |
100VA |
Voltage adjustment method |
Digital setting setting resolution 0.01 kV |
voltmeter |
True effective value display |
Digital |
AC0.00 kV to 5.00 kV |
Accuracy |
± 1.5% rdg (1000 V or more) |
± 15 V (1000 V or less) |
Waveform |
Sinusoidal distortion factor 5% or less (at 40 Ω load 5 kV output) |
Output frequency |
50 Hz / 60 Hz ± 0.2% |
Switch setting is possible regardless of the input power frequency. |
Current detection unit |
Current measurement range |
0.01 mA to 20 mA (2 range configuration) |
ammeter |
True effective value display (digital) |
Measurement range and resolution |
0.00 mA to 9.99 mA, 0.01 mA (10 mA range) |
0.00 mA to 20.0 mA, 0.1 mA (20 mA range) |
Measurement accuracy |
± 2% rdg ± 0.05 mA (10 mA range) |
Ą 2% rdg ± 0.5 mA (20 mA range) |
(Note) rdg (reading value, display value, indicated value) |
The value currently being displayed, the value currently indicated by the measured value. |
Insulation resistance test section · Test voltage · Measurement range |
Rated voltage |
DC 500/1000 V |
No-load voltage |
1 to 1.2 times the rated voltage |
Rated measuring current |
1 mA to 1.2 mA |
Short circuit current |
4 mA to 5 mA (500 V) 2 mA to 3 mA (1000 V) |
Measurement range |
0.5 MΩ to 999 MΩ (500 V) / ± 4% rdg |
Accuracy |
1 MΩ to 999 MΩ (1000 V) / ± 4% rdg |
1000 MΩ to 2000 MΩ / ± 8% rdg |
Measurement resistance range |
2 MΩ, 20 MΩ, 200 MΩ, 2000 MΩ (500 V), 4 MΩ, 40 MΩ, 400 MΩ, 2000 MΩ (1000 V) |