General specification |
Type name |
TS-EB 0266 |
Input power supply |
Single phase AC 100 V ± 10% 50/60 Hz |
Dimension / mass |
W320 × H150 × D380 (rubber feet 20 mm excluding projections) Approximately 20 Kg |
Test specification |
Test voltage |
AC 0 to 2.5 kV One end grounded type |
Transformer capacity |
500 VA (rated for 30 minutes) |
Voltage setting method |
Set voltage range 0.00 to 2.50 kV (0.01 kV increments) Automatic boosting by slider
motor drive up to set voltage by test start Voltage setting by booster
(variable range: 15 to 90 sec / fs) |
Frequency / waveform |
Input power synchronization |
voltmeter |
Digital AC 0.0 to 2.50 kV (± 1.5% fs) Analog AC 0 to 5 kV (± 5% fs) |
ammeter |
0.01 mA to 200 mA |
Timer setting |
Off, 0.5 to 99.9 sec (0.1 sec step), 100 to 999 sec (1 sec step) |
Test judgment method |
Window · Comparator type (digital setting)
UPPER-FAIL judgment (when measured current value exceeds set upper limit value) and
LOWER-FAIL judgment (measured current value is less than set lower limit value) |
Current setting range |
0.1 to 9.9 mA (0.1 mA step), 10 to 200 mA (1 mA step) |
Function · Interface |
Memory function |
Up to 10 types of test voltage and judgment value can be registered |
Hold function |
It holds (displays) the measured value at the end of the test |
External control |
External I / O, RS-232C, status out (patrol lighting I / F) |