Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Cảm biến đo áp suất CV3100 Labom

Mã sản phẩm: CV3100
Sử dụng cho Model:
Hãng SX: Labom
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng
Features

Modular pressure transmitter with internal or flush mounted diaphragm.
Output signal: 4…20 mA, HART® protocol (rev. 6), option, PROFIBUS PA, option
Function modules: Multifunctional display with 5-segment digital display and bar graph, Switching module with 2 flating channels, maximum 0.5 A switching current, electrically isolated at all sides, without additional auxiliary power
Function module replacement on site without recalibration "plug and measure"
Limits of measuring range 0...80 mbar to 0...400 bar

Yêu Cầu Giá Tốt

Thông tin sản phẩm

Material wetted parts
Sensor: Piezoresistive Thin film
Sensor diaphragm: 1.4404/1.4435 (316L) 1.4542 (630)
Socket: 1.4404/1.4435 (316L) 1.4301/1.4404 (304/316L)
Gasket: EPDM-FDA listed --
Measuring system
Sensor: Piezoresistive Thin film
Sensor filling: Synthetic oil, free of silicone FD1, FDA listed without
Accuracy
Limit point setting: per DIN 16086
Reference conditions: per EN 60770-1
Linearity errors: ≤ 0.15 % of span
TD 5:1 no modification
Hysteresis: ≤ 0.05 % of nominal range
Repeatability: ≤ 0.05 % of nominal range
Calibration position: Vertical mounting position
Influence of mounting position: ≤ 3,5 mbar
Long-term drift (EN 60770-1): ≤ 0.1 % / year of nominal range
Temperature effect: Lower range value / upper range value
  Range 0…60 °C: ± 0.15 % / 10K of nominal range
  Range < 0 °C, > 60 °C: ± 0.2 % / 10K of nominal range
Output
General: Delay time: approx. 160 ms
  Measuring cycle: 6 measurements / second
  Measuring range
setting:
Turndown 5:1
Basic module: 4…20 mA Signal: 4…20 mA, 2-wire
  Current range: 3,8…20,8 mA
  Current limitation: approx. 22 mA
  Alarm state: < 3.6 mA, optional > 21 mA
  Damping: 0…120 seconds
  Load RB: RB ≤ (UV-12V DC)/0,022A [Ohm]
UV = supply voltage
Basic module: PROFIBUS PA Signal: digital per IEC 61158-2
  Protocol: EN 50170-PROFIBUS PA, Profile 3.0
  Sensor address: 0…126 (126 = factory setting)
  Power consumption: constantly 11 mA
  Fault current IFDE : 2 mA
  Damping: 0…300 seconds
  Parameterization: SIMATIC PDM

Bình luận