- Safe and convenient to operate: Full micro-computerized devices, built-in imported high-performance CPU, with high reliability, according the prompt of the interface to set test values, you need not touch the tested equipment, the equipment automatically finish the test, the test personnel far from high voltage circuit ,to ensure their personal safety.
- Full-featured: CT volt-ampere characteristic test, inflection point calculation and error curve, CT ratio polarity test, contrast calculation, secondary side circuit inspection, PT volt-ampere characteristic test, PT ratio polarity test. Automatically records show that all kinds of test results, and can save the print test records, real-time instrument comes with the clock function, greatly convenient for the user to document the test report.
- High output capacity: the maximum output voltage for volt-ampere characteristic tests up to 1000V, external booster (optional) can be maximum increased to 2500V, variable ratio test and a flow-through test maximum current 600A,the output capacity of equipment is 5KVA, can do the volt-ampere characteristic test of 500kV,1A current transformers
- Large-screen 320 × 240 dot matrix character graphical interface, directly display volt-ampere characteristic curve after the test, graphics are clear, aesthetic, easy to analyze, equipped with micro-printer, can print the curves and test results at any time.
- Built-in Flash memory device can store large amounts of test data, which will not be lost after power off. When the filed tests finished, people can check and print the tests results indoors.
- U-disk dump: support U disk to dump test data, through PC to read, easy to operate.
- PC with powerful software, beautiful outlook, easy to operate, can save and print the test results, convenient to manage the test data.
- Apparatus can achieve AC220V and AC380V adaptive power supply. When the AC power input terminals connected to AC220V voltage, AC voltage output is 0 ~ 600 V, when the input terminals to AC380V voltage, AC voltage output is 0 ~ 1000 V. Users can choose according to actual needs.
Thông số kỹ thuật
Power supply
|
AC380V
|
AC220V
|
CT volt-ampere characteristic test
|
stand-alone
|
0~1000V、0~20A
|
0~600V、0~20A
|
external booster(optional)
|
0~2500V、0~3A
|
0~2000V、0~3A
|
Resolution
|
0.1V、1mA
|
Accuracy
|
≤0.5%
|
5% or 10% error curve
|
Accuracy
|
≤1.0%
|
CT ratio test
|
Range
|
5~25000A/5A(5000A/1A)
|
Accuracy
|
≤1.0%
|
PT volt-ampere characteristic test
|
stand-alone
|
0~1000V、0~20A
|
0~600V、0~20A
|
external booster(optional)
|
0~2500V、0~3A
|
0~2000V、0~3A
|
Resolution
|
0.1V、1mA
|
Accuracy
|
≤0.5%
|
PT ratio polarity test
|
Range
|
1~10000
|
Accuracy
|
≤1.0%
|
Primary current injecting
|
Current Range
|
0~600A
|
Current injecting time
|
5 min
|
Power supply voltage
|
AC220V or AC380V 50HZ
|
Working environment temperature
|
-10℃ ~ +50℃
|
host volume
|
420×330×320㎜3
|
host weight
|
39kg
|
external booster (optional)volume
|
310×310×330㎜3
|
external booster (optional)weight
|
25kg
|