- High-strength aluminum alloy shell body, the circuit equipped with phase locked loop and switched capacitor filter, based on which the instrument play an effective role in shielding power frequency and radio frequency interference.
- DC/AC converter can transform DC current into AC low-frequency constant current for the measurement.
- Auxiliary grounding resistance can change in the range of 0 ~ 1.8KΩ (RC), 0 ~ 40KΩ (RP), that will not affect measurement results.
- Without manual adjustment to zero, with three and a half bits LCD display, the instrument can be used for measuring the grounding resistance, the conductor resistance of low resistance, soil resistivity and AC ground voltage. If tested loop encounter failure, the screen displays "1" indicating large range, which consistent with conventional measurement practices.
Thông số kỹ thuật
Operating Conditions
|
Temperature
|
0℃~+45℃
|
Relative humidity
|
≤85%RH
|
Measuring range and constant current value (virtual value)
|
resistance
|
0~2Ω(10mA),2~20Ω(10mA),20~200Ω(1mA)
|
voltage
|
AC0~19.99V
|
Measurement accuracy and resolution
|
accuracy
|
0~0.2Ω≤±3%±1d
|
0.2Ω~199.9Ω≤±1.5%±1d
|
1~19.99V≤±3%±1d
|
resolution
|
0.001Ω、0.01Ω、0.1Ω、0.01V
|
Measurement errors caused by auxiliary grounding resistance and ground voltage
|
Auxiliary grounding resistance
|
RC (between C1 and C2)
|
0~2Ω,2~20Ω≤1KΩ
|
20~200Ω≤2KΩ
|
RP (between P1 and P2) <40KΩ error ≤ ± 5%
|
Ground voltage (frequency virtual value)
|
≤ 5V error ≤ ± 5%
|
Power supply and power dissipation
|
Maximum power dissipation
|
≤2W
|
DC
|
8×1.5V(AA,R6) battery
|
AC
|
220V/50Hz
|
Dimensions and weight
|
Dimension
|
220×200×105mm3
|
Weight
|
≤1.4kg
|