- High precision: the core components are imported devices, high integration, high precision, good repeatability, high reliability.
- Highly intelligent: the calibration process without the constant temperature room, in any effective temperature range can test the SF6 gas density relay, pressure gauge. Pressure and temperature dynamic compensation algorithm is used to automatically complete the pressure measurement and 20℃ equivalent, completely solve the problem of SF6 gas density relay field calibration difficult.
- Powerful: can test a set of alarm signals, two groups of blocking signals, and has the normal temperature pressure gauge and 20 degree pressure gauge calibration function.
- Friendly man-machine interface: the configuration of large LCD screen and the rotation mouse human-machine interface, easy to operate, beautiful interface, all parameters and status at a glance. Test process automatically completed, without manual intervention, to avoid tedious manual gas path adjustment operation.
- Transition joints: portable instrument, equipped with a variety of types of transition joints, most types of switch density relay calibration site without disassembly, it is the best choice for SF6 density relay calibration.
- Less gas consumption: gas consumption is very small, without waste SF6 gas, test costs are low, no pollution.
- Data storage: built-in memory, real-time storage test results, with power down data saved function, check and print the history of test data at any time.
- Report printing: report forms print test result, and carries on the intelligence analysis to the test result.
- Clock modification: with the function of online modification of the system clock.
- U disk to save: support U disk to save the test data, read through the PC, easy and fast operation
- PC software powerful: easy to operate, to achieve data analysis, save, print and generate test report, convenient for data centralized management
Thông số kỹ thuật
Working voltage
|
AC220V, 50Hz
|
Power
|
50W
|
Accuracy
|
0.2
|
Pressure measurement range
|
0~1MPa
|
Pressure measurement resolution
|
0.001MPa
|
Temperature measurement range
|
-20℃~80℃
|
Temperature measurement resolution
|
0.1℃
|
Humidity
|
≤90%RH
|
Test objects
|
Single alarm, single atresia, single alarm & single atresia, single alarm & dual atresia
|
SF6 density display mode
|
The pressure in the test environment, the equivalent pressure at 20℃
|
Display
|
320×240 LCD
|
Operation
|
Rotating mouse
|
Printer
|
High-speed micro printer
|
Print method:
|
Report type
|
Storage Capacity
|
Save 50 test results at the same time
|
Dimension
|
Host
|
400×320×160mm
|
Weight
|
6.2kg
|
Accessory
|
400×370×200mm
|
|