Design: |
High quality case with bayonet ring per EN 837-1 S1, material: stainless steel mat.-no.-Nr. 1.4301 (304) ; with blow-out device, material: PUR, ventilation valve, material: PUR |
|
Alternative:
Safety design with blow-out back and solid baffle wall per DIN EN 837-1 S3, Material: Stainless steel 1.4301 (304), pressure compensation diaphragm: silicone |
Nominal size: |
DN 100 or DN 160 |
Degree of protection per EN 60529: |
- Without filling: IP 65
- With filling S1 case: IP 65
- With filling S3 case: IP 66 |
Case filling: |
Labofin |
Atmosph. pressure compensation: |
Standard case, filled, ≤ 10 bar: with ventilation valve
< 10 bar filling plug, material: PUR
Safety case, filled: with pressure compensation diaphragm, material: silicone |
Case seal: |
Material gasket: NBR |
Window: |
Non splintering laminated glass |
Contact lock: |
Stainless steel with NBR gasket, sealable |
Measuring element: |
Bourdon tube
< 60 bar: c-type
≥ 60 bar: spiral |
Movement: |
Stainless steel segment |
Scale: |
Pure aluminium, white with black inscrip-tion |
Pointer: |
Pure aluminium, black, with micro adjust-ment for zero point correction |
Mounting: |
Via process connection. Optional with flange for surface mounting or for flush mounting with DIN mounting flange. |
Electronical connection: |
Connection plug with cable gland M20 x 1.5 and removable test cover, material: Macrolon |
Electrical evaluation unit: |
The following evaluation units conform to the requirements of the Association of Technical Inspection Agencies (VdTÜV) Fact Sheet No. 100. |
|
Pepperl+Fuchs, switching amplifier:
- Typ KHA6-SH-Ex1, PTB 00 ATEX 2043
- Typ KFD2-SH-Ex1, PTB 00 ATEX 2042 |
|
The use of alternative evaluation units is within the responsibility of the operator. |
|
The data sheets of the electrical evalua-tion units are to be observed. |
Weight: |
- NS 100 without filling: approx. 0.9 kg
- NS 160 without filling: approx. 1.8 kg
- NS 100 with filling: approx. 1.5 kg
- NS 160 with filling: approx. 3.6 kg |