Features
Comprises twin 150mm focal length monochromators
Built in a single casting, for optimised rigidity and robustness
Stepping motor and sine bar driven diffraction gratings
Predictable sinusoidal wavelength accuracy
One entrance and up to two exit ports with automated mirror-based port selection for use of multiple input/ output configurations
Control electronics integrated and fully automated via USB 2.0 interface
Opto-Mechanical
Wavelength scanning |
By integrated stepping motor and micro-stepping drive electronics |
Slit type |
Fixed, micrometer or motorised variable |
Slit width & height |
10µm-10mm (W) x 20mm (H) |
Filter size |
- 6 position filter wheel : 25mm diameter
- 8 position filter wheel : 22mm diameter |
Optical Performance (quoted for typical 2400 g/mm gratings)
Automation
Computer interface |
USB 2.0 |
Software control |
BenWin+/SDK |
Spectral range |
200-600nm |
Linear dispersion |
5.4nm/mm |
Wavelength accuracy |
± 0.15nm, ± 0.05nm with software correction |
Wavelength reproducibility |
± 0.05nm |
Resolution full/reduced slit height |
0.25/0.1nm |
Stray light rejection at 2.5 FWHM |
10-8 |
Wavelength sweep speed |
30nm/s |
Configuration |
Pair symmetrical Czerny Turner monochromators, additive dispersion |
Focal length |
Total focal length: 300mm
Individual Units: 150mm |
Grating mount |
Kinematic |
Number of gratings |
1 pair |
Grating size |
33x33mm |
Aperture ratio |
f/4 |
Number of entrance/exit ports |
1 entrance, 1-2 exits |
Spectral range |
Within range 200nm-2400nm |
|