Features
Houses standards in light tight enclosure with easy access to position devices under test
Colour shifted (2856K quartz tungsten halogen (QTH) standards (CIE illuminant A)
Output level of the illuminance calibrator is varied via four neutral density filters, providing five calibration levels
Lab jack and height gauge, featuring a laser alignment aid, allow accurate positioning of device under test
Output level of the luminance calibrator is varied via three neutral density filters, providing four calibration levels
Includes high performance 610 programmable constant current power supply
Calibration traceable to Physicalish Technische Bundesanstalt (PTB, Germany).
LuxCal250
Lamp type |
Quartz Tungsten Halogen lamp, G6.35 base |
Nominal lamp power and voltage |
250W, 12V |
Operating current |
10.4A (adjusted to obtain 2856K operation) |
CCT |
2856±10K |
Expected lifetime |
2000 hours |
Calibration frequency |
100 hours use/ one year recommended |
Dimensions |
600L x 500W x 700H (mm) |
Typical illuminance level one (no filter) |
19000 lx |
Typical illuminance level two (ND1) |
1800 lx |
Typical illuminance level three (ND2) |
180 lx |
Typical illuminance level four (ND3) |
20 lx |
Typical illuminance level five (ND4) |
2 lx |
LumiCal50
Lamp type |
Grit-blasted Quartz Tungsten Halogen lamp, G6.35 base |
Nominal lamp power and voltage |
50W, 12V |
Operating current |
4.0A (adjusted to obtain 2856K operation) |
Expected lifetime |
2000 hours |
Calibration frequency |
100 hours use/ one year recommended |
Typical luminance level one (no filter) |
21000 cd.m-2 |
Typical luminance level two (ND1) |
1800 cd.m-2 |
Typical luminance level three (ND2) |
200 cd.m-2 |
Typical luminance level four (ND3) |
30 cd.m-2 |
Typical luminance level five (ND4) |
3 cd.m-2 |
|