Goniometer
Angular rotation (no dead angle) |
C-γ mode
Horizontal rotational axis (C-planes) = 360°
Vertical rotational axis (γ) = 0-140/160°
B-β mode
Horizontal rotational axis (B-planes) = 360°
Vertical rotational axis (β )= 140/160° |
Angular resolution |
Horizontal rotational axis: 0.05°
Vertical rotational axis: 0.05° |
Area occupied by goniometer |
2m diameter circular area |
Maximum sample size |
C-γ mode: 500 mm diameter, 300mm deep
B-β mode: 1.65m long, 500mm diameter |
Maximum sample weight |
C-γ mode: 20 kg
B-β mode: 40 kg |
Photometric detector
Spectral function |
CIE 1924 V(λ) Spectral Luminous Efficiency Function for Photopic Vision |
Spectral response range |
380-780nm |
Photopic match, f1 |
<3% |
Angular response, f2 |
<1.5% |
Linearity, f3 |
<0.2% |
Typical responsivity |
5x 10-10A/ lux |
Calibration traceability |
Physikalisch-Technische Bundesanstalt (PTB, Germany) |
Diffuser diameter |
7mm |
Stray light suppression |
Variable aperture baffle tube |
Connector |
BNC |
Amplifier gain ranges |
1010-105 V/A |
Maximum input |
100µA |
Input impedance |
Virtual ground |
Gain accuracy & stability |
±1 %, 200 ppm/ °C |
Output stability |
5-500 ppm/ °C dependent on gain range |
ADC resolution |
4 ½ digit BCD (0 to 19999) (>14 bit resolution) |
Luminous intensity range |
0.01D2-200000D2 (D photometric distance [m]) |
Area occupied by photometer frame |
0.5 x 0.5m |
General
Power supply |
PMC MAC- 110/220V 50/60Hz, ORM400– transformer |
Interface |
PMC MAC & ORM400- USB 2.0 |
|