Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Fibre Spectral Attenuation Spectrometer FSL300 Bentham

Mã sản phẩm: FSL300
Sử dụng cho Model:
Hãng SX: Bentham
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

The FSL300 system provides accurate assessment of fibre spectral loss, paramount in quality control processes, material research and optical fibre development. Employing the internationally recognised reference test method, the cut-back technique, the FSL300 provides high measurement precision across all fibre types.

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

Features


Accurately assesses fibre spectral loss
Employs internationally recognised cut-back technique
The wavelength band may be extended throughout UV-vis and into the IR beyond 5000nm
Fully-automated via PC, responsible for dB/km loss calculation as a function of wavelength 
High-precision assessment of a large range of fibre types, including silica, doped glass, plastic and photonic bandgap fibres
Optical stability is ensured by the use of a quart tungsten halogen light source driven by a constant-current power supply
Light source and launch optics are bolted firmly to the monochromator to negate mechanical instability over time

FSL300_Si

Spectral range 350-1100nm
Typical spectral data interval 5nm
Typical spectral bandwidth (FWHM) 5nm
Light source Optically chopped IL1
Monochromator TMc300 with 1200 g/mm gratings
Detector 2.4x 2.4mm room temperature Silicon: 350-1100nm
Wavelength accuracy GA-T312R0U5 ± 0.2nm

FSL300_InGaAs

Spectral range 800-1700nm
Typical spectral data interval 5nm
Typical spectral bandwidth (FWHM) 5nm
Light source Optically chopped IL1
Monochromator TMc300 with 1200 g/mm gratings
Detector 1mm dia. room temperature InGaAs: 800-1700nm
Wavelength accuracy GA-T30831U2: ± 0.3nm

FSL300_InGaAs_EX

Spectral range 800-2500nm
Typical spectral data interval 5nm
Typical spectral bandwidth (FWHM) 5nm
Light source Optically chopped IL1
Monochromator TMc300 with 830 & 600 g/mm gratings
Detector 1mm dia. thermoelectrically cooled extended InGaAs: 800-2500nm
Wavelength accuracy GA-T30831U2: ± 0.3nm
GA-T306R1U6: ± 0.4nm

FSL300_PbSe

Spectral range 1500-5000nm
Typical spectral data interval 10nm
Typical spectral bandwidth (FWHM) 10nm
Light source ILD-QH-IR-CHOP
Monochromator TMc300 with 600 & 300 g/mm gratings
Detector 3mm dia. thermoelectrically cooled PbSe: 1000-5000nm
Wavelength accuracy GA-T306R1U6: ± 0.4nm 
GA-T303R3U0: ± 0.8nm


Bình luận