Parameter |
Range |
Accuracy |
Resolution |
Flue Gas Measurements |
Oxygen |
0-25 % |
± 0.2 % Vol abs |
0.1% Vol |
Carbon Monoxide |
0-4000 ppm |
± 10 ppm or ± 5 % of reading |
1 ppm |
Efficiencies( net, gross, condensing)1 |
0 to 120 % |
|
0.1 % |
Air Ratio1 |
1 to 9.99 |
|
0.01 |
CO/CO2 Ratio1 |
0-9999 |
|
0.0001 |
Excess Air1 |
0 to 99.9% |
|
0.1% |
Heat Losses1 |
0 to 99.9 % |
|
0.1 % |
CO2%1 |
0 to CO2% max |
± 0.3% Vol |
0.1% |
Temperature and Pressure measurements |
Gas Temperature |
-50 to 1200°C |
± 2°C |
0.1 °C |
Air Temperature |
-50 to 100°C |
± 1°C |
0.1 °C |
Draught |
± 50 mbar |
±0.02 mbar |
0.01 mbar |
Differential Pressure |
± 300 mbar |
±0.02 mbar |
0.01 mbar |
Pre-Programmed fuels |
Nat Gas, Oil Heavy, Oil Light, Propane, Butane
Wood@0%,Wood@10%,Wood@20%, Wood@30%, Wood@40%, Wood@50%, Pellets
Kerosene, Coal, Diesel, Petrol, Coke, Anthracite
4 user fuels |
Memory, communications and data storage |
Memory |
1000 data set |
|
|
Storage medium |
Internal memory and Micro SD card ( card not included) |
Data export |
Trough USB or SD card in CSV format |
Communications |
Micro USB, Micro SD Card, IRDA and optional Bluettooth® |
General |
|
|
|
Display |
4 lines touch screen colour display |
Power |
Standard Rechargeable, Li-Ion batteries 2250 mA/h, USB charging AC 100-240 VAC, 47-63 Hz – Wall charger and cable included |
Battery Life |
10 hours |
Weight |
390 grams |
Dimension(W x Lx H |
82x169x44 mm |
IP Protection |
IP 40 |
Operating Temperature |
5 to 45 °C |
Storage Temperature |
-20 to 50 °C |
Warranty |
full 2 years manufacturing warranty on instruments and cells |
5 years warranty |
5 years extended warranty when annual serviced by EIUK |
Approval |
Conform to BS 7967 and BS EN 50379-2 |
Flue Gas Probe |
250 mm stainless steel shaft, integrated temperature and pressure measurement, probe stop with handle and 2.7 meter hose with large water trap with filter |
Notes |
1 calculated value |