| General specification |
| Type name |
TS-EB 0265 |
| Input power supply |
Single phase AC 100 V ± 10% 50/60 Hz |
| Dimension / mass |
W250 × H150 × D330 (rubber feet 20 mm excluding projections) Approximately 11 Kg |
| Test specification |
| frequency |
50 to 400 Hz (set by 1 Hz step by thumb rotary switch) |
| Waveform |
Sinusoidal distortion factor 1.0% or less |
| Output voltage |
AC 0 to 400 V (Note 1)
Manual setting with one end grounding type voltage adjustment volume (linearly variable) |
| Blocking current setting |
0.1 mA to 100 mA (Note 1) |
| Output Capacity |
200VA |
| Timer setting |
Off, 0.5 to 99.9 sec (0.1 sec step), 100 to 999 sec (1 sec step) |
| Test judgment method |
Window · Comparator method
UPPER-FAIL judgment (when the measured current value exceeds the set upper limit value) and
LOWER-FAIL judgment (when the measured current value is less than the set lower limit value) |
| voltmeter |
Digital AC 0.0 to 2.50 kV (± 1.5% fs) |
| ammeter |
Digital 0.01 mA to 100 mA (± 3% fs ± 20 μA)
|
| Current measurement resolution |
0.01 mA (2 mA / 8.0 mA range)
0.1
mA (32 mA range) 1 mA (100 mA range)
|
| Function · Interface |
| Memory function |
Up to 10 types of test voltage and judgment value can be registered |
| Hold function |
It holds (displays) the measured value at the end of the test |
| External control |
External I / O, RS-232C, status out (patrol lighting I / F) |