Working condition
|
-15℃~40℃ RH<80%
|
Power
|
AC220V±10% Frequency Unlimited
|
AC high voltage output
|
400V~2200V ±2% interval 100V 50VA
|
DC high voltage output
|
200V~600V ±2% interval 50V
|
Temperature control induction furnace
|
50VA;output current:25mA
|
Temperature control induction furnace
|
Maximum power 500W
|
Temperature control range
|
<100℃
|
Temperature control error
|
±0.5℃
|
Temperature measurementresolution
|
1℃
|
Temperature control time
|
Room temperature to 90 ℃, less than 20min
|
Measuring range
|
tgδ
|
Unlimited
|
Cx
|
15PF~300PF
|
Rx
|
10M-20T
|
Resolution
|
△tgδ
|
0.001%
|
△Cx
|
0.01pF
|
△Rx
|
0.01
|
Accuracy
|
△tgδ
|
±(reading×0.5%+0.020%)
|
△Cx
|
±(reading×0.5%+0.5PF)
|
△Rx
|
±reading×10%
|
Relative dielectric constant
|
εr
|
automatic calculation according to Cx
|
Volume resistivity
|
ρ
|
automatic calculation according to Rx
|
Dimension
|
450×310×360mm3
|
Weight
|
18kg
|
Distance between HV andLV
|
2mm
|
Capacitance of hollow cup
|
60±2PF
|
The maximum measurement voltage
|
Power frequency 2000V
|
Dielectric loss of hollow cup
|
tgδ<1×10-4
|
Liquid capacity
|
About 40ml
|
Electrode material
|
stainless steel
|
Dimension
|
70mm(D)×120mm(H)
|