The ELEMENTRAC OH-p determines oxygen and hydrogen in inorganic samples by inert gas fusion in an impulse furnace with temperatures in excess of 3,000 °C. The ELEMENTRAC OH-p guarantees precise and fast sample analysis. The analyzer covers a wide range of applications such as metal,ceramics and other inorganic materials.


Yêu Cầu Giá Tốt

Thông tin sản phẩm

Application Examples


alloys, cast iron, ceramics, copper, refractory metals, steel, ...


Product Advantages


  • simultaneous oxygen/hydrogen determination with inert gas fusion technique
  • NEW: closed gas management and optimized gas circulation for sensitive OH determination
  • NEW: enlarged Schuetze Reagent tube for precise oxygen measurements
  • NEW: gas flow system with electronic gas flow control and new leakage test
  • NEW: water-cooled sample port system for effective removal of atmospheric gases
  • flexible configurations and measuring ranges for O and H
  • high sensitivity IR and TC cells with low detection limits
  • short analysis time
  • rapid, precise, accurate and reliable element determination
  • powerful 8,5 kW* impulse furnace for temperatures in excess of 3,000 °C
  • economic analysis of grains without capsules
  • NEW: chemicals and tubes are hidden behind a door (removable)
  • NEW: powerful software supporting data and application export, with comment fields
  • single and multipoint calibration (linear regression)
  • NEW: cooling via tap water, heat exchanger or chiller
  • New design allows operation in production control and laboratory


Features


Measured elements hydrogen, oxygen
Samples inorganic
Furnace alignment vertical
Sample carrier graphite crucibles
Field of application ceramics, engineering / electronics, steel / metallurgy
Furnace electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C
Detection method solid state infrared absorption for oxygen, thermal conductivity for hydrogen
Typical analysis time 120 - 180 s
Chemicals required magnesium perchlorate, Schuetze reagents, sodium hydroxide
Gas required compressed air, nitrogen 99.995 % pure, all gases with (2 - 4 bar / 30 - 60 psi)
Power requirements 3~ 400 V, 50/60 Hz, max. 8,500 W
Dimensions (W x H x D) 57 x 77 x 63 cm
Weight ~ 161 kg
Required equipment balance (resolution 0.0001g), monitor, PC
Optional accessories carrier gas purification, external chiller, gas calibration unit
- * limited to 6.8 kw in application settings