Application Examples
alloys, cast iron, ceramics, copper, refractory metals, steel, ...
Product Advantages
- simultaneous oxygen/nitrogen or oxygen/hydrogen determination with inert gas fusion technique
- NEW: closed gas management and optimized gas circulation for sensitive ONH determination
- NEW: use of cost efficient argon as carrier gas possible
- NEW: powerful catalyst furnace for precise oxygen measurement
- NEW: gas flow system with electronic gas flow control and new leakage test
- NEW: water-cooled sample port system for effective removal of atmospheric gases
- flexible configurations and measuring ranges for O, N and H
- high sensitivity IR and TC cells with low detection limits
- short analysis time
- powerful 8,5 kW* impulse furnace for temperatures in excess of 3,000 °C
- economic analysis of grains without capsules
- NEW: chemicals and tubes are hidden behind a door (removable)
- NEW: powerful software (supporting data and application export, comment fields, and many more)
- single and multipoint calibration (linear regression)
- NEW: cooling via tap water or heat exchanger or chiller
- New design allows operation in production control and laboratory
Features
Measured elements |
hydrogen, nitrogen, oxygen |
Samples |
inorganic |
Furnace alignment |
vertical |
Sample carrier |
graphite crucibles |
Field of application |
ceramics, engineering / electronics, steel / metallurgy |
Furnace |
electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C |
Detection method |
solid state infrared absorption for oxygen thermal conductivity for nitrogen and hydrogen |
Typical analysis time |
120 - 180 s |
Chemicals required |
copper oxide, magnesium perchlorate, Schuetze reagents, sodium hydroxide |
Gas required |
compressed air, helium 99.995 % pure, nitrogen 99.995% pure, argon 99.995% pure (if required), all gases with (2 - 4 bar / 30 - 60 psi) |
Power requirements |
3~ 400 V, 50/60 Hz, max. 8,500 W |
Dimensions (W x H x D) |
57 x 77 x 63 cm |
Weight |
~ 161 kg |
Required equipment |
balance (resolution 0.0001g), monitor, PC |
Optional accessories |
carrier gas purification, external chiller, gas calibration unit |
- |
* limited to 6.8 kw in application settings |
|