The PhotoTest450 is a trolley-based high irradiance source of UVB-UVA– and visible monochromatic radiation, designed for in vivo application, representing a key tool in phototoxicity testing, photogenotoxicity testing and in the investigation of potential phototoxic risk in trials of topical and systemic drugs.

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

Features


Efficiently coupled 450W xenon source/ monochromator tuneable light source
Provides coverage of UV and visible (280-600nm)
Accurate exposure time delivered by use of automated shutters 
Flexible liquid light guide beam delivery
Accurate real time dosimetry of monochromatic probe using calibrated integrating sphere based power meter
System calibration traceable to PTB, Germany
User-friendly Windows TES software fully automates process with minimal training required (USB 2.0)
No lamp water-cooling required



Applications



Phototoxicity testing
Photogenotoxicity testing 
Investigating potential phototoxic risk in trials of topical and systemic drugs

Tuneable light source

Lamp rating 450W
Typical lamp current 27A
Lamp lifetime 2000h
Current sensing Current transducer coupled to ORM400
Monochromator configuration Asymmetric, single Czerny-Turner
Focal length 300mm
Grating mount and drive Stepping motor driven triple grating turret
Diffraction gratings installed 2400 groove/mm holographic, 250nm blaze
1200 groove/mm ruled, 400nm blaze
1200 groove/mm ruled, 500nm blaze
Grating size 2400 g/mm– 68x84mm
1200 g/mm—90x90mm
Aperture ratio f/3.6 (at all grating angles)
Number of entrance/exit ports 1 entrance, 1 exit
Slit type Internal motorised variable
Slit width & height 10µm-10mm (W) x 20mm (H)
Order sorting 6-position filter wheel (includes shutter), 25mm diameter filters
Spectral range 280-600nm
Maximum bandwidth 2400 g/mm– 13.5nm
1200 g/mm—27nm
Wavelength accuracy ± 0.1nm (2400 g/mm ); ± 0.2nm (1200 g/mm )
Wavelength reproducibility ± 0.025nm (2400 g/mm ); ± 0.05nm (1200 g/mm )

Beam Dosimetry

Integrating sphere diameter and coating 4”, BaSO4
Sphere detector Silicon photodiode
Sphere calibration Spectral Responsivity, A.W-1.nm-1
Calibration traceability PTB, Germany
Peak responsivity (typical) 0.14mA.W-1
Photocurrent measurement Six-decade trans-impedance amplifier/ ADC