- The RH Systems Model 973-A Dew Point Mirror was specifically designed for measurement of humidity in gas insulated switchgear (GIS) systems. Humidity measurement data is displayed in PPMv, PPMw and dew/frost point at either GIS pressure or standard pressure. Based on the chilled mirror principle, the 973-A ensures direct and accurate humidity measurement with minimal gas loss.
- Full Color Touch Screen
- Easy, Automated Usage
- Verifiable Calibration
Thông số kỹ thuật
Measuring range: |
|
Frost/Dew Point |
-55 ...+20°C at atmospheric pressure |
PPMv Range |
10 ...20,000 |
Accuracy: |
|
Frost/Dew point |
±0.5°C |
PPMv |
±6% of reading |
Reproducibility: |
|
Frost/Dew Point |
± 0.2°C |
PPMv |
± 3% of reading |
Measuring Mode: |
Automatic, with gas shutoff |
Thermoelectric Mirror cooling: |
3-stage, 80°C depression |
Cooling: |
Air |
Display: |
Active Matrix Color Graphic LCD |
Digital I/O: |
RS-232 |
Gas couplings: |
Quick connect fitting |
Mirror temperature sensor: |
PRT-100 |
Sample gas flow rate: |
0.5 l/min |
Sample gas circuit: |
PTFE (Teflon) and Stainless Steel |
Operation temperature: |
-10 ...+40°C |
Storage temperature: |
-10 ...+40°C |
Ambient humidity: |
Instrument 90% RH max., non condensing in use |
Voltage: |
100 ...240 VAC, 50/60 Hz |
Power consumption: |
200 Watt |
Width: |
440 mm (600 mm with transport case) |
Height: |
155 mm (310 mm with transport case |
Depth: |
410 mm (450 mm with transport case) |
Weight: |
13 kg (23 kg with transport case) |
|