Features
Run a small current on the test object and test it.
The current can be selected from 2 fixed values and volume setting, and the voltage can be varied.
Combined with a special Box, it can be tested safely in a short time.
Soft starter is considered to prevent excessive voltage and current from being applied.
At the end of the test, it is safe to discharge the test object through the resistance.
Specification
Power supply |
1 φ 100 V ± 10%, 50/60 Hz, input capacity 500 VA or less |
Output specification |
format |
One end grounded type, positive polarity |
Constant current |
DC fixed 100 μA, 1 mA and VR variable (~ 1 mA) |
Constant voltage |
DC 1.5 kV to 15 kV, variable VR |
Meter |
voltmeter |
4 digits, Accuracy: 1.5% |
ammeter |
3 1/2 digit, accuracy: 2.5%, with current monitor terminal on the back side |
protection |
Cutoff current |
DC 2 mA error ± 5% |
Overheating |
When internal temperature reaches 85 ° C, output stop |
Safety function |
○ Open the door of the test box and stop the
test ○ With discharge function of the test item |
Outline |
Body: W300 × D350 × H210 (mm) Painting color: 2.8 Y 7.7 / 0.3 Test Box: W 500 × D 319 × H 324
(mm) |
mass |
Body: 13 kg, Test Box: 10 kg |
|