Thiết bị đo điện trở cách điện 
Model :MIS-A ANALOG 
Hãng sản xuất : Multi 
    
        
            | Insulation resistance | 
         
    
    
        
            | Rated voltage and effective measuring range | 
            MIS-1A:50V-10MΩ,125V-20MΩ,250V-50MΩ 
            MIS-2A:125V-20MΩ,250V-50MΩ,500V-100MΩ 
            MIS-3A:125V-20MΩ,250V-50MΩ,1000V-2000MΩ 
            MIS-4A:250V-50MΩ,500V-100MΩ,1000V-2000MΩ | 
         
        
            | Center scale | 
            MIS-1A:0.2MΩ/0.5MΩ/1MΩ 
            MIS-2A:0.5MΩ/1MΩ/50MΩ 
            MIS-3A:0.5MΩ/1MΩ/2MΩ 
            MIS-4A:1MΩ/2MΩ/50MΩ | 
         
        
            | Minimum measurable resistance at rated voltage | 
            MIS-1A:0.05MΩ,0.125MΩ,0.25MΩ 
            MIS-2A:0.125MΩ,0.25MΩ,0.5MΩ 
            MIS-3A:0.125MΩ,0.25MΩ,1MΩ 
            MIS-4A:0.25MΩ,0.5MΩ,1MΩ | 
         
        
            | Rated current | 
            1mA+20%-0% | 
         
        
            | Open circuit voltage | 
            Rated voltage+30%-0% | 
         
        
            | Short circuit current | 
            <2mA | 
         
    
 
    
        
            | SPECIFICATIONS | 
         
    
    
        
            | Withstanding Voltage | 
            AC 3700V, 1 minute (Between input terminal and outer case) | 
         
        
            | Storage Temperature | 
            -10℃~60℃, 80% RH max (Without condensation) | 
         
        
            | Operating Temperature | 
            0℃ to 40℃, 80% RH max. (Without condensation) | 
         
        
            | Power supply | 
            1.5V (AM-3, R6 or "AA" size)×6 | 
         
        
            | Weight | 
            Approx. 330g (Excluding batteries) | 
         
        
            | Safety standard | 
            IEC 61010-1 (2001) 
            Installation categoryⅡ, 600V phase to earth. | 
         
        
            | Size | 
            170(W)×105(D)×54(H)mm | 
         
        
            | Accessories | 
            Line test lead:1 
            Earth test lead:1 
            Batteries:6 
            Test lead case:1 
            Belt:1 
            Instruction manual:1 | 
         
        
            | Optional accessory | 
            Test probe with remote switch | 
         
        
            | E.M.C. standard | 
            EN 61326 | 
         
        
            | Function | 
            Insulation resistance, AC voltage, Battery check | 
         
        
            | Meter movement | 
            100μA, 870Ω, Taut band meter | 
         
        
            | Constructional standard | 
            In accordance with IEC1557-2 or JIS C1302 (1994) | 
         
        
            | Overload protection | 
            120% of the highest nominal output voltage (10 sec.) | 
         
        
            | Battery check | 
            DC 6.3V~9.5V | 
         
        
            | Low battery limit | 
            DC 6.3V | 
         
        
            | Temperature characteristics(0~40℃) | 
            ±5% rdg of specified accuracy | 
         
        
            | Accuracy(DC 50V) | 
            First effective range:0.01MΩ~5MΩ ±5%rdg、10MΩ~50MΩ ±30%rdg 
            Second effective range: | 
         
        
            | Accuracy(DC 125V) | 
            First effective range:0.02MΩ~10MΩ ±5%rdg 
            Second effective range:0.01MΩ~0.02MΩ 10MΩ~20MΩ ±10%rdg,10MΩ~100MΩ ±30%rdg | 
         
        
            | Accuracy(DC 250V) | 
            First effective range:0.05MΩ~20MΩ ±5%rdg 
            Second effective range:0.02MΩ~0.05MΩ 20MΩ~50MΩ ±10%rdg,50MΩ~100MΩ ±30%rdg | 
         
        
            | Accuracy(DC 500V) | 
            First effective range:0.1MΩ~50MΩ ±5%rdg 
            Second effective range:0.05MΩ~0.1MΩ 50MΩ~100MΩ ±10%rdg | 
         
        
            | Accuracy(DC 1000V) | 
            First effective range:2MΩ~1000MΩ ±5%rdg 
            Second effective range:1MΩ~2MΩ 1000MΩ~2000MΩ ±10%rdg | 
         
        
            | AC voltage (50/60Hz) | 
            Range/Accuracy/Input impedance/Maximum input voltage 
            AC 600V/±2.5% of full scale/Approx. 1.5M/AC 600V rms | 
         
    
 
 
  
                             |