• 3 test voltages: 100 V - 250 V - 500 V
• Guard terminal
• Programmable timer
• Pass-fail test
• Low resistance and continuity measurement
• AC/DC Voltmeter up to 400 V - True RMS
• Auto-range
• Rechargeable battery
Nominal test voltages
100 V - 250 V - 500 V. Negative, DC voltage.
Resistance measurement range
From 10 kΩ up to 200,000 MΩ.
Short circuit current
1.5 ± 0.3 mA.
Test Voltage Accuracy
-0 / +15% for resistances between 5 MΩ and open circuit.
Basic Accuracy
R ≤ 50 GΩ: ±5% of reading ± 2 digits.
R > 50 GΩ: ±10% of reading ± 2 digits.
Low resistance measurement and continuity tester
Range: 0.05 Ω to 150 Ω.
Accuracy: ±5% of reading ± 2 digits.
Test current: up to 200 mA.
Continuity indicator: Activates at R < 5 Ω ± 0.5 Ω.
Built-in timer
5 - 10 - 15 - 30 seconds.
1 - 2 - 3 - 5 - 10 - 15 minutes.
Pass / fail test
0,2 - 0,3 - 0,4 - 0,5 - 0,6 - 0,7 - 0,8 - 0,9 MΩ.
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - 10 - 20 - 30 - 40 - 50 - 60 - 70 - 80 - 90 - 100 - 200 - 300 - 400 - 500 - 600 - 700 - 800 - 900 MΩ.
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - 10 GΩ.
Voltmeter
True RMS.
Range: 0 - 400 V CA/CC.
Accuracy: ±3% of reading ± 2 digits.