Inspection Technology
|
Eddy Current.
|
Operating Frequencies
|
60 kHz, 120 kHz, 240 kHz, 480 kHz. Option of 960 kHz.
|
Conductivity Range
|
0.5 % IACS to 110 % IACS, 0.28-64 MS/m
|
Accuracy
|
Probe must be in thermal equilibrium with test material, and instrument and probe stabilized prior to testing Minimum 15-minute warm up time for stabilisation. At 20°C: 0-20% IACS: ± 0.05 IACS At 20°C: >20% IACS ± 0.25% of readout
|
Display Resolution
|
Up to 3 decimal places
|
Lift Off
|
13 mm probe compensated to 0.020” (0.5mm) 7 mm probe compensated to 0.010” (0.25 mm)
|
Temperature Measurement
|
In-probe sensor (accurate to 0.5 °C). Range 0 °C to + 50 °C
|
Automatic Temperature Compensation
|
Conductivity measurements are corrected to the 20°C value.
|
Environmental Range
|
0 to 95% relative humidity, 0°C to + 50°C for reliable operation
|
Display
|
2.8” (70mm) 320 x 240 pixels colour display. LCD with selectable backlight.
|
Construction & Storage
|
High impact, splash-proof, moulded UL94-5VA flame-retardant ABS case. Protective rubber boot to protect the unit, probes, probe cable, operator manual on USB, and removable stand.
|
Conductivity Standards
|
On top of unit. Removable for value verification, and when attached ensures thermal equilibrium.
|
Power
|
2 x 1.5 V AA NiMH Batteries, Approx up to 9 hrs life. Recommended rechargeable batteries: Panasonic Eneloop Pro, 1.2 V min.2500 mAh or equivalent) Can also use non-rechargeable AA cells
|
Size
|
163mm (6,4”) x 80mm (3.1”) x 25mm (1.0”)
|
Weight
|
350g (0.77 pounds) including batteries
|
Data Logger Memory
|
Removable 2GB micro SD Card allowing over 1 million readings to be stored.
|
PC Connectivity
|
USB port for charger and PC communications
|
Probes
|
13mm diameter for 60 kHz to 480 kHz 8mm probe operates at 60 kHz to 480 kHz Probes are interchangeable with simple operator resetting procedure. Probes are field exchangeable and do not require return to manufacturer for calibration.
|
Accessories
|
A range of conductivity reference standards traceable to US and European standards are available for in field use. Up to five can be mounted on an aluminium anodised Setting Reference Blocks Holder (Part number: ASIG003)
|