Model |
High load FGPX-H Series |
FGPX-250H |
FGPX-500H |
Rated capacity (RC) |
± 2500N |
± 5000N |
± 250Kg |
± 500Kg |
Display range |
± 2500N |
± 5000N |
± 250.0Kg |
± 500.0Kg |
Display resolution |
1N / 0.1Kg |
|
Measurement system |
Normal measurement, plus peak measurement, minus peak measurement |
Display cycle |
1 times / sec, 2 times / sec, 3 times / sec., 5 times / sec, 10 times / sec, 20 times / sec. |
Sampling period |
1000 times / sec. |
accuracy |
± 0.2% RC and ± 1/2 digit (at 23 ℃) |
Temperature effect |
Gain: ± 0.02% / LOAD / ℃ , zero: ± 0.02 Pasento / RC / ℃
(zero point drift can be canceled by the tracking function) |
display |
LCD with code 4-digit display (character height 12mm) |
Judgment LED (Hi_NG, OK, Lo_NG) 3 pieces |
Overload capability |
150% RC |
Comparator function |
Hi_NG, OK, determine the Lo_NG, output in the determination LED and the output signal |
communication |
USB |
Yes (can communicate with the PC by a dedicated communications software. Connection cable is standard accessory) |
RS-232C |
Yes (can communicate with the PC by a dedicated communication command. Connection cable is optional) |
Digimatic |
It can be connected to a Mitutoyo Digimatic printer DP-1VR |
analog |
± 1V (output to ± display range), accuracy ± 50mV, 12bit D / A converter system |
Output update 1000 times / sec zero can be set, load resistance 10kΩ or more |
output |
overload |
PUSH, PULL open collector output (maximum DC30V / 5mA) |
comparator |
HI, OK, LO open collector output (maximum DC30V / 5mA) |
input signal |
ZERO (external zero setting), HOLD (external hold) |
Power supply |
AC adapter (DC9V / 200mA) or built-in nickel-metal hydride battery (rechargeable) |
During the measurement can be battery operable time charge: Mitsuru about 8 hours after charging |
Charging time: up to 16 hours (automatically end when it comes to full charge) |
Auto power-off function |
10 minutes (when the AC adapter connection is possible continuous use), invalid settings are also possible |
Memory function
|
Continuous memory 1000 / alone memory 100 cases / standard memory 50 items
(3 kinds of memory mode switchable)
Yes statistics function (maximum, minimum, peak, average, standard deviation) |
Operating temperature limit |
0 ~ 40 ℃ (except that there is no condensation) |
Operating humidity range |
35 ~ 85% RH (but no condensation) |
External dimensions |
Width 82 × thickness 57 × length 214mm |
mass |
About 900g |
About 1000g |
accessories |
Measurement adapter (6 types), the hanger, AC adapter, carrying case, USB cable, handle |
Communications software |
Dedicated communication software free download from (USB compatible) Homepage |