The F6000 is a new addition to Factair’s range of Safe-Air Testers. This instrument enables breathing-air tests to be easily and quickly carried out without the need for any chemical reagent tubes, making the unit ideal for use in environments where glass products are not permitted. The unit supports a range of international standards and in the UK this ensures complete compliance with the relevant requirements of COSHH L5.
The breathing-air supply is subjected to an automatically controlled test against the requirements of EN12021 2014, or the alternative international standards availabe from the instrument’s menu. The F6000 incorporates electronic cells for carbon monoxide, carbon dioxide, and oxygen. Moisture levels within the breathing air are measured by a dewpoint sensor, which is specially suited to sample air quality testing, ensuring an accurate moisture reading within the standard test time. The instrument then displays the result in pressure or atmospheric dewpoint levels, as well as providing the concentration in mg/m3 . It also includes an electronic flowmeter for measuring air flow rates up to 600 l/min, a digital display for airline pressure and ambient temperature reading.