General Parameters
|
Display
|
5.7” TFT touch screen
|
Operating Power
|
AC 110 V / 220 V, ±10%,50/60 Hz
|
Fuse
|
6 A
|
Max Power
|
200 W
|
Storage Space
|
Unlimited (PC)
|
Communication Method
|
Ethernet, RJ45
|
Errors Detection
|
Errors will be shown on LCD screen and ongoing operation will be stopped.
|
Operating Status
|
LED indicator, LCD display
|
Power Supply Connection to EUT
|
Cable with 4 mm2 banana plug
|
Output Measuring Method
|
Via standard 1 m co-axial cable
|
HV Output Measuring Method
|
Via HV cable with 4 mm2 banana plug
|
Dimension
|
19” / 6U
|
Weight
|
About 30 Kg
|
Operating Temperature
|
15 – 35℃
|
Operating Humidity
|
45%–75%
|
Operating Air Pressure
|
86 kPa–106 kPa
|
Technical Parameters
|
Combination Wave
|
Test Voltage
|
0.3 kV – 6 kV (±10%)
|
Communication Wave
|
Test Current
|
7.5 A - 150/200 A (±10%)
|
Voltage Waveform
|
Front time: 10 μs ±30%,
Half-value time: 700 μs ±20%
|
Current Waveform
|
Front time: 5 μs ±20%;
Half-value time: 320 μs ±20%
|
Source Impedance
|
15 ohm /40 ohm (automatic switching)
|
Output Polarity
|
+, -, +/-
|
Pulse Period
|
6s-99s (the shortest time is subject to the test voltage)
|
No. of Impulses
|
1–999
|
Triggering Mode
|
Automatic, Manual, External Triggering
|