Measured medium |
Alcohols, milk beverage, seasoning, pure water, chemical or other media that will not corrode SUS316L. |
Pressure type |
Positive, compound, and negative pressures can be measured. |
Pressure range |
Can be specified within the range of 50 kPa to 3 MPa (Calibration:25kPa~3MPa) of rated capacity.
Specify more than 50% of rated capacity as the calibration range. |
Output type |
4~20mA (2 wire system) |
Specification 1 (Sensor assembly)
Pressure Port Type/Size |
ISO Ferrule, VARIVENT®, ISO Cap nut, Screw mount, Other special orders |
Rated Capacity(R.C.) |
50, 100, 200, 300, 500, 1000kPa, 3MPa |
Material |
Sensor: SUS316L |
Finish(Sensor Part) |
Electro-Chemical Buffing |
Safe Overload Rating |
400%R.C. (Rated capacity 3MPa: 150%R.C.) |
Destruction Point |
1000%R.C. (Rated capacity 3MPa: 300%R.C.) |
Specification 2 (Amplifier + Sensor assembly)
Accuracy (*1)
(Non Linearity, hysteresis and repeatability included.) |
±0.3%R.C. (Rating: 300, 500, 1000kPa, 3MPa)
±0.35%R.C. (Rating: 50, 100, 200kPa) |
Compensated Temperature Range (Wetted part) |
High temperature: -5~150°C |
Safe Temperature Range (Wetted part) |
High temperature: -5~150°C |
Temperature Effect on Zero Balance |
±0.35%R.C./10°C (Rating: 300, 500, 1000kPa, 3MPa)
±0.3%R.C./10°C (Rating: 50, 100, 200kPa) |
Temperature Effect on Load |
±0.35%R.C./10°C (Rating: 300, 500, 1000kPa, 3MPa)
±0.3%R.C./10°C (Rating: 50, 100, 200kPa) |
Ambient Temperature and Humidity |
Temperature:0~50°C(No freezing), Humidity:0~90%RH(No condensation) |
Material |
SUS304(Case material), SUS304(Radiating fin) |
IP Rating |
Equivalent to IP65 |
Weight |
Approx. 700g |
Accessories |
Operation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy)
1 of hexagon nut(Cap nut only) |
|