■Measurement of the high vacuum area is possible.
0~1500Pa(abs), 0~5kPa(abs) and 0~10kPa(abs)
■High leak performance
■Maximum allowable pressure (400kPa)
| Measured medium |
Water, oil, gases or other media that will not corrode Ceramic + SUS316L + O-ring. |
| Pressure type |
Compound pressure can be measured. |
| Measured Range |
Can be specified within the range of 1500Pa(abs), 5kPa(abs) and 10kPa(abs) of calibration. |
| Output type |
Specify among 2 to 20mA(4 wire system), 1 to 5V |
Specification
| Non Linearity |
±0.8%F.S. or less |
| Hysteresis |
±0.8%F.S. or less |
| Range of Use Temperature |
0~70°C (No condensation or freezing) |
| Temperature Effect on Load |
±0.5%F.S. / ±10°C or less |
| Safe Overload Rating |
±0.5%F.S. / ±10°C or less |
| Connection Screw Material |
SUS316L |
| Sensor Element |
Alumina ceramic |
| O-ring Material |
Selected from EPDM or Fluororubber (Class 4D) or Perfluoro-Elastomer |
| Case Material |
SUS304 |
| IP Rating |
Equivalent to IP54 |
| Power |
DC24V (±10%) |
| Mass |
Approx. 470g |
| Accessories |
Operation manual (1 copy), Test report (1 copy) |
|