■Suitable for pressure measurement of ultrapure water and corrosive media such as strong acids or alkalis.
■Integrally molded PFA body offers high leak-proof performance.
■Employs a leak indicator to alert you in case gas permeation occurs.
Measured medium |
Ultrapure water, strong acid, strong alkali or other media that will not corrode PFA. |
Pressure type |
Positive, compound and negative pressures can be measured. |
Measured Range |
Rated capacity : 0(-0.1)~0.4, 0.6, 1MPa
Specify more than 60% of rated capacity as the calibration range(F.S.).
*(Example)
Rated capacity : In the case of 0.4MPa
0~0.2MPa (=0.2MPaF.S.)
→ Manufacture is impossible.
-0.1~0.2MPa (=0.3MPaF.S.)
→ Manufacture is possible. |
Output type |
Specify among 4 to 20 mA (2 wire system), 1 to 5 V, 0 to 5 V, 0 to 10 V. |
Specification
Accuracy |
±1%R.C.(25°C)(Non Linearity, Hysteresis, Repeatability included) |
Gas Permeability Detection |
Leak indicator lamp changes from yellow to red when the medium is
acidic, and from yellow to blue if alkaline.
Exchange time: Becomes red or blue |
Compensated Temperature Range
(Temperature of medium to be measured) |
15~40°C(No condensation or freezing) |
Safe Temperature Range
(Temperature of medium to be measured) |
0~80°C(No condensation or freezing) |
Temperature Effect |
±0.5%R.C.(15~40°C) |
Safe Overload Rating |
0.4→0.6MPa, 0.6→0.9MPa, 1→1.25MPa |
Pressure Port Material |
PFA fluoroplastic |
Housing Material |
PP (Polypropylene) |
Leakage |
3×10-8Pa・m³/sec. or less |
Sealing |
Seal-less due to integral molding |
Screw type |
1/4, 3/8, 1/2, 3/4Tube, Flared, Screw-in |
Cable |
PVC coated 4-core (current type is 2-core) shielded
3m pigtail(UL-compliant) |
IP Rating |
Equivalent to IP55 |
Weight |
Approx. 360g(Cable 3m included) |
Accessories |
Operation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy) |
|