■5 analog outputs and 4 BCD outputs can be selected and used together.
■VPC and sensor are calibrated as a pair, including the sensor cable length, and constitute a high-accuracy digital pressure meter.
■Fast response analog output without microcomputer is available that is suitable to fluid pressure measurement.
Can be combined with sensors |
Can be combined with all models of pressure sensors and load cells. |
Sensor specification |
Refer to the pages of pressure sensors and load cells. |
Calibration of amplifier and sensor |
VPMC amplifiers and pressure sensors or load cells have been calibrated as a pair.
Amplifiers and sensors have been calibrated as a pair under the same serial number. |
Specification
Display |
1999 3½ digit Super luminosity Red LED(Character height: Approx. 15mm) |
Indication Conversion Rate |
3 times/sec.(Standard) Can be set in the range of 1 to 15 times/s. |
Accuracy |
±0.05%F.S.±1digit |
Use Temperature Range |
0~55°C(No freezing) |
Use Humidity Range |
0~90%RH(No condensation) |
Dimensions |
φ104×56(D:Protrusions not included) |
Weight |
Approx. 650g(Varies with specification) |
Accessories |
Operation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy)(*1) |