- The LCMK-EM is a low-resistance, low-noise probe tip holder for the MM3A-EM micromanipulator. It enhances your system's nanoprobing capability by allowing low-capacity, low-current measurements on conductive samples.
- Standard tungsten probe tips can be inserted into the front of the LCMK-EM tip holder. Triax cabling and feedthroughs carry signals through to a connector which fits all conventional parameter analyzers.
- The plug-in tool can be quickly and easily fitted to your existing MM3A-EM micromanipulator.
- Current measurement limit: 3 fA @ 1 Hz
- Insulation leakage current: 50 fA at 1 V
- Signal conductor resistance: 0.9 Ω
- Maximum voltage: ±210 V
- Maximum current: 200 mA
- Maximum frequency: 100 MHz