Iwatsu Model |
CS-3100 |
CS-3200 |
CS-3300 |
Collector supply |
|
Mode |
High voltage |
AC, ± Rectified SIN, ± DC, ± LEAKAGE |
Large current |
- |
Pulse only |
|
Maximum peak power |
120 mW, 1.2 W, 12 W, 120 W, 390 W
*390 W can be selected excluding the maximum peak voltage
setting at 3,000 V range.
|
- |
Large current mode (400 W, 4 kW) |
Large current mode (400 W, 4 kW, 10 kW) |
|
High voltage mode |
Maximum peak current
(maximum peak pulse current) |
Maximum peak voltage |
75 mA (150 mA) |
3,000 V (2,500 V for AC) |
750 mA (1.5 A) |
300 V |
7.5 A (15 A) |
30 V |
|
|
|
|
Loop correction |
The stray capacitance between the collector terminal and the ground of the fixture is compensated by the hardware in the high voltage mode. Digital compensation is also available. |
|
Large current mode (pulse only) |
- |
Maximum peak current |
Maximum peak voltage |
400 A |
40 V |
40 A |
40 V |
|
|
|
Maximum peak current |
Maximum peak voltage |
1,000 A |
40 V |
400 A |
40 V |
40 A |
40 V |
|
|
|
|
Pulse width/measurement point |
- |
The pulse width is variable in the range from 50μs to 400 μs.
Measurement points can be specified (with a resolution of 10 μs/step). |
|
Maximum number of data points |
20 to 1,000 points can be specified per trace. |
|
MEASUREMENT MODE |
REPEAT, STOP/SINGLE, SWEEP |
Step generator |
|
Current mode |
Amplitude range: 50 nA to 200 mA, 21 ranges, at 1-2-5 step
Maximum current: ±2A / Offset: ±10 times STEP AMPLITUDE setting |
|
Voltage mode |
Amplitude range: 50 mV to 2 V, 6 ranges, at 1-2-5 step
Maximum voltage:±40V / Offset: ±10 times STEP AMPLITUDE setting |
|
Step rate |
Staircase wave: 2times of 50 Hz or 60 Hz
(50 Hz or 60 Hz in the AC mode) |
Staircase wave: 2times of 50 Hz or 60 Hz (50 Hz or 60 Hz in the AC mode)
Pulse: The pulse changes in the range from 80ms to 1,000 ms.
(The lowest frequency limited by the maximum peak power consumption setting.) |
|
Pulse step |
The pulse width is variable in the range from 50 μs to 400 μs at 10μs resolution. |
|
Number of steps |
0 to 20 steps |
AUX output |
|
Range |
OFF, -40V to +40 V, variable with at 100mV resolution |
Vertical axis |
|
Collector current |
Range |
High voltage mode: 1 μA/div to 2 A/div, 20 ranges, at 1-2-5 step. |
- |
High current mode: 100 mA/div to 50 A/div, 9 renges, at 1-2-5 step. |
High current mode: 100 mA/div to 100 A/div, 10 renges, at 1-2-5 step.
|
|
Emitter current (LEAKAGE) |
Range |
1 nA/div - 2 mA/div, 20 ranges,at 1-2-5 step |
Horizontal axis |
|
Collector voltage |
Range |
High voltage mode: 50 mV/div to 500 V/div, 13 ranges, at 1-2-5 step |
- |
High current mode: 50 mV/div to 5 V/div, 7 ranges, at 1-2-5 switching |
|
Base/emitter voltage |
Range |
50 mV/div to 5 V/div, 7 ranges, at 1-2-5 step |
Miscellaneous |
|
Display |
8.4-inch color TFT-LCD (SVGA 800 x 600 pixels) |
|
Data save/recall |
Internal: Memory (setup: 256, REF waveform: 4)
External: Removable storage connected to the USB port
(setup, waveform save / recall,
screen hard copy of display(BMP, TIFF, PING) |
|
USB |
1 port (USB1.1) Waveform Save / Recall, Hard copy of display(BMP, TIFF, PING) |
|
Remote control |
Remote control by LAN : 1 port (100BASE-TX) |
|
Power supply |
AC 100 V - AC 240 V, single phase, 50/60 Hz, Approx. 500VA |
Mechanism section |
|
Dimensions (mm) |
424 (W) x 555.2 (L) x 221 (H) (without the external projection portion) |
424 (W) x 555.2 (L) x 354.5 (H)
(without the external projection portion) |
|
Weight |
Approx. 30 kg (without optional accessories) |
Approx. 45 kg (without optional accessories) |