Hotline: 0984.843.683 Email: info@ttech.vn  Zalo: 0984.843.683

Thiết bị hiệu chuẩn MK-300K Soukou

Mã sản phẩm: MK-300K
Sử dụng cho Model:
Hãng SX: Soukou
Xuất xứ
Bảo hành
Tình trạng

Tình trạng: Còn hàng

Đặt hàng

It can be calibrated equipment together various meters such board in one

Liên Hệ Hỗ Trợ

    • SĐT: 0984 843 683

    • Mail: info@ttech.vn

      • Zalo: 0984.843.683

Thông tin sản phẩm

Specification


Use power Single-phase AC100V ± 10V 50 / 60Hz
Output voltage Three-phase AC0 ~ 150 / 300V 150VA (continuous)
Output current Three-phase AC0 ~ 5A 150VA (continuous)
Power conditioning Batch control, phase adjustment (single-phase / three-phase switching)
Frequency adjustment 50Hz / 60Hz / continuously variable 
* variable range 40 ~ 70Hz (0.01Hz unit) continuously variable by the rotary encoder
Power factor (phase) adjustment Coarse / variable 
* coarse range +0.5 / +0.8 / 1.0 / -0.8 / -0.5 step by the switch variable 
* variable range + 0 ~ 1.0 ~ -0 (0 ~ 360 ° phase) continuously by the rotary encoder variable
Display voltmeter (V) Digital power meter
Phase voltage ± 0.25% rdg + 0.05% FS (resolution 0.01V)
Line voltage ± 0.75% rdg + 0.1% FS (resolution 0.01V)
Ammeter (A) ± 0.25% rdg + 0.05% FS (resolution 0.001A)
Frequency meter (Hz) ± 0.01% rdg (resolution 0.01Hz)
Reactive power meter (kVAR) More than 5% of ± 1.5% rdg FS (resolution 0.001kVAR)
Reactive power meter (kVARh) ± 1.5% rdg (resolution 0.001kVARh)
Power-factor meter ± 1.0% rdg (Resolution 0.01)
Power meter (kWh) 
power meter (kW)
Power factor ± 0.8 or more and 1.0 or less current 2% or more and 100% or less than ± 1.0% rdg of FS 
power factor ± 0.8 or more and 1.0 or less current 1% 2% or more less than ± 1.5% rdg of FS 
power factor of 0.5 or more 0.8 less than the advance or delay 
more than 20% to 100% ± 1.0% rdg of current FS 
power factor of 0.5 or more 0.8 less than the advance or delay of 
less than 20% ± 1.5% more than 2% of the current FS rdg
Output Protection Voltage, if the current output of the overload condition, activate the protective function, the output will stop. At this time, the alarm lamp lights up.
Test items Voltmeter, ammeter, power meter, power meter, reactive power meter, reactive power meter, power factor meter, calibration of the frequency meter
usage environment Temperature / 5 ~ 40 ℃ humidity / 20% to 80%
External dimensions 250 (D) × 540 (W) × 520 (H)
weight About 30kg
accessories Test code set (with storage bag) owner manual case cover

Bình luận