Use power |
AC100V ± 10% 50 / 60Hz |
Output voltage |
~ 3KV AC0.1
DC0.05 ~ 3KV (positive polarity) |
Output current |
AC, DC both 165mA (30 minute rating) |
Overcurrent detection operation |
To cancel the test state at about 110% of each rated output |
Output system |
AC output is a combination of regulated power supply unit and the high-voltage transformer
DC output, the voltage doubler rectification by the switching system |
Voltmeter
measurement range
resolution
measurement accuracy |
Digital meter (7-segment LED)
0 ~ 3KV
0.001KV
AC ± 1.0 Pasento rdg ± 30Dgt (50V or more)
DC ± 0.5 Pasento rdg ± 10Dgt (more than 50V) |
Ammeter
measurement range
resolution
measurement accuracy |
Digital meter (7-segment LED)
0 ~ 20 / 200MA
0.001 / 0.01MA
AC ± 1.0 Pasento rdg ± 30Dgt (5MA or more)
DC ± 0.5 Pasento rdg ± 10Dgt (more than 0.2mA) |
Voltage adjustment method |
Manual control |
Timer over |
1s ~ 99m55s digital timer over |
Discharge function |
In the case of DC voltage output, to discharge the residual charge in the test OFF (built-in discharge resistance 250Keiomega)
* AC voltage output does not work |
External dimensions |
392 (D) × 506 (W) × 420 (H) |
weight |
About 31kg |
accessories |
Test code (with storage bag) Operating Instructions |