Oscilloscope – vertical (analog) |
9300 models |
9300-25 models |
Channels |
PicoScope 9341: 4
All other models: 2 |
Bandwidth, full |
20 GHz |
25 GHz |
Bandwidth, narrow |
10 GHz |
12 GHz |
Pulse response rise time, full bandwidth |
17.5 ps (10% to 90%, calculated) |
14.0 ps (10% to 90%, calculated) |
Pulse response rise time, narrow bandwidth |
35.0 ps (10% to 90%, calculated) |
29.2 ps (10% to 90%, calculated) |
Noise, full bandwidth |
< 1.5 mV RMS typical, < 2.0 mV RMS maximum |
< 1.9 mV RMS typical, < 2.5 mV RMS maximum |
Noise, narrow bandwidth |
< 0.8 mV RMS typical, < 1.1 mV RMS maximum |
< 1.0 mV RMS typical, < 1.3 mV RMS maximum |
Input connectors |
2.92 mm (K) female, compatible with SMA and PC3.5 |
Resolution |
16 bits, 40 μV/LSB |
Scale factors (sensitivity) |
1 mV/div to 500 mV/div in 1-2-5 sequence with 0.5% fine increments |
Operating input voltage |
1 V p-p within ±1 V range (with digital feedback, single-valued)
±400 mV relative to channel offset (without digital feedback, multi-valued) |
Nominal input impedance |
(50 ±1) Ω |
Accuracy |
±2% of full scale ±2 mV over nominal temperature range (assuming temperature-related calibrations are performed) |
DC offset range |
Adjustable from −1.000 V to 1.000 V in 10 mV increments (coarse). Also adjustable in fine increments of 0.2 mV. Referenced to the center of display graticule. |
Maximum safe input voltage |
16 dBm, or ±2 V (DC + peak AC) |
Optical/electrical converter (PicoScope 9321) |
Bandwidth (−3 dB) |
9.5 GHz typical |
Effective wavelength range |
750 nm to 1650 nm |
Calibrated wavelengths |
850 nm (MM), 1310 nm (MM/SM), 1550 nm (SM) |
Transition time |
51 ps typical (10% to 90% calculated from tR = 0.48/optical BW) |
Noise |
4 μW (1310 & 1550 nm), 6 μW (850 nm) maximum @ full electrical bandwidth |
DC accuracy |
±25 μW ±10% of full scale |
Maximum input peak power |
+7 dBm (1310 nm) |
Fiber input |
Single-mode (SM) or multi-mode (MM) |
Fiber input connector |
FC/PC |
Input return loss |
SM: −24 dB typical
MM: −16 dB typical, −14 dB maximum |
Timebase (sequential equivalent time sampling mode) |
Digitizing rate |
With digital feedback (single-valued): DC to 1 MHz
Without digital feedback (multi-valued): DC to 40 kHz |
Delta time interval accuracy |
Sequential equivalent time |
For >200 ps/div: ±0.2% of of delta time interval ±12 ps. |
Random equivalent time |
For =200 ps/div: ±5% of Delta time interval ±5 ps, whichever is smaller. |
Real time |
±0.2% of of delta time interval or full horizontal scale, whichever is greater. |
Scale |
Sequential equivalent time |
5 ps/div to 3.2 ms/div (main, intensified and delayed) |
Random equivalent time |
50 ns/div to 100 us/div |
Real time |
2 μs/div to 100 ms/div |
Roll |
200 ms/div to 10 s/div |
Time interval resolution |
Sequential equivalent time |
=(screen width) / (record length) or approximately 64 fs, whichever is larger. |
Random equivalent time |
4 ns min. |
Real time |
1 μs min. |
Data record length |
32 to 32 768 points (single channel) in x2 sequence |
Deskew |
1 ps resolution, 100 ns max |
Acquisition modes |
Sample (normal), average, envelope |
|