| Oscilloscope – vertical (analog) | 
            9300 models | 
            9300-25 models | 
         
    
    
        
            | Channels | 
            PicoScope 9341: 4 
            All other models: 2 | 
         
        
            | Bandwidth, full | 
            20 GHz | 
            25 GHz | 
         
        
            | Bandwidth, narrow | 
            10 GHz | 
            12 GHz | 
         
        
            | Pulse response rise time, full bandwidth | 
            17.5 ps (10% to 90%, calculated) | 
            14.0 ps (10% to 90%, calculated) | 
         
        
            | Pulse response rise time, narrow bandwidth | 
            35.0 ps (10% to 90%, calculated) | 
            29.2 ps (10% to 90%, calculated) | 
         
        
            | Noise, full bandwidth | 
            < 1.5 mV RMS typical, < 2.0 mV RMS maximum | 
            < 1.9 mV RMS typical, < 2.5 mV RMS maximum | 
         
        
            | Noise, narrow bandwidth | 
            < 0.8 mV RMS typical, < 1.1 mV RMS maximum | 
            < 1.0 mV RMS typical, < 1.3 mV RMS maximum | 
         
        
            | Input connectors | 
            2.92 mm (K) female, compatible with SMA and PC3.5 | 
         
        
            | Resolution | 
            16 bits, 40 μV/LSB | 
         
        
            | Scale factors (sensitivity) | 
            1 mV/div to 500 mV/div in 1-2-5 sequence with 0.5% fine increments | 
         
        
            | Operating input voltage | 
            1 V p-p within ±1 V range (with digital feedback, single-valued) 
            ±400 mV relative to channel offset (without digital feedback, multi-valued) | 
         
        
            | Nominal input impedance | 
            (50 ±1) Ω | 
         
        
            | Accuracy | 
            ±2% of full scale ±2 mV over nominal temperature range (assuming temperature-related calibrations are performed) | 
         
        
            | DC offset range | 
            Adjustable from −1.000 V to 1.000 V in 10 mV increments (coarse). Also adjustable in fine increments of 0.2 mV. Referenced to the center of display graticule. | 
         
        
            | Maximum safe input voltage | 
            16 dBm, or ±2 V (DC + peak AC) | 
         
    
 
    
    
        
            | Optical/electrical converter (PicoScope 9321) | 
         
    
    
        
            | Bandwidth (−3 dB) | 
            9.5 GHz typical | 
         
        
            | Effective wavelength range | 
            750 nm to 1650 nm | 
         
        
            | Calibrated wavelengths | 
            850 nm (MM), 1310 nm (MM/SM), 1550 nm (SM) | 
         
        
            | Transition time | 
            51 ps typical (10% to 90% calculated from tR = 0.48/optical BW) | 
         
        
            | Noise | 
            4 μW (1310 & 1550 nm), 6 μW (850 nm) maximum @ full electrical bandwidth | 
         
        
            | DC accuracy | 
            ±25 μW ±10% of full scale | 
         
        
            | Maximum input peak power | 
            +7 dBm (1310 nm) | 
         
        
            | Fiber input | 
            Single-mode (SM) or multi-mode (MM) | 
         
        
            | Fiber input connector | 
            FC/PC | 
         
        
            | Input return loss | 
            SM: −24 dB typical 
            MM: −16 dB typical, −14 dB maximum | 
         
    
 
    
    
        
            | Timebase (sequential equivalent time sampling mode) | 
         
    
    
        
            | Digitizing rate | 
            With digital feedback (single-valued): DC to 1 MHz 
            Without digital feedback (multi-valued): DC to 40 kHz | 
         
        
            | Delta time interval accuracy | 
         
        
            |   Sequential equivalent time | 
            For >200 ps/div: ±0.2% of of delta time interval ±12 ps. | 
         
        
            |   Random equivalent time | 
            For =200 ps/div: ±5% of Delta time interval ±5 ps, whichever is smaller. | 
         
        
            |   Real time | 
            ±0.2% of of delta time interval or full horizontal scale, whichever is greater. | 
         
        
            | Scale | 
         
        
            |   Sequential equivalent time | 
            5 ps/div to 3.2 ms/div (main, intensified and delayed) | 
         
        
            |   Random equivalent time | 
            50 ns/div to 100 us/div | 
         
        
            |   Real time | 
            2 μs/div to 100 ms/div | 
         
        
            |   Roll | 
            200 ms/div to 10 s/div | 
         
        
            | Time interval resolution | 
         
        
            |   Sequential equivalent time | 
            =(screen width) / (record length) or approximately 64 fs, whichever is larger. | 
         
        
            |   Random equivalent time | 
            4 ns min. | 
         
        
            |   Real time | 
            1 μs min. | 
         
        
            | Data record length | 
            32 to 32 768 points (single channel) in x2 sequence | 
         
        
            | Deskew | 
            1 ps resolution, 100 ns max | 
         
        
            | Acquisition modes | 
            Sample (normal), average, envelope | 
         
    
  
                             |