Oscilloscope specifications |
Specifications with PicoConnect 442 1000 V CAT III probe |
Input channels |
4 channels |
4 channels |
Analog bandwidth (–3 dB) |
20 MHz with D9 to BNC adaptors
15 MHz with PicoConnect 441 probe |
10 MHz |
Rise time (calculated) |
17.5 ns with D9 to BNC adaptors
23 ns with PicoConnect 441 probe |
35 ns |
Bandwidth limit |
100 kHz or 1 MHz (selectable) |
100 kHz or 1 MHz (selectable) |
Vertical resolution, 12-bit mode |
12 bits on most input ranges
11 bits on ±10 mV range |
12 bits |
Vertical resolution, 14-bit mode |
14 bits on most input ranges
13 bits on ±20 mV range
12 bits on ±10 mV range |
14 bits |
Enhanced vertical resolution
12-bit mode |
16 bits on most input ranges
15 bits on ±10 mV range |
16 bits |
Enhanced vertical resolution
14-bit mode |
18 bits on most input ranges
17 bits on ±20 mV range
16 bits on ±10 mV range |
18 bits |
Input type |
Differential
9-pin D-subminiature, female |
Differential
9-pin D-subminiature, female |
Input characteristics |
1 MΩ ±1%, in parallel with 17.5 pF ±1 pF (each differential input to ground).
< 1 pF difference between ranges. |
16.7 MΩ ±1%, in parallel with 9.3 pF ±1 pF (each differential input to ground) |
Input coupling |
AC/DC |
AC/DC |
Input sensitivity
(10 vertical divisions) |
2 mV/div to 10 V/div |
±0.5 V/div to ±200 V/div |
Input ranges (full scale) |
±10 mV, ±20 mV, ±50 mV, ±100 mV, ±200 mV, ±500 mV,
±1 V, ±2 V, ±5 V, ±10 V, ±20 V, ±50 V |
±2.5 V, ±5 V, ±12.5 V, ±25 V, ±50 V, ±125 V, ±250 V, ±500 V, ±1000 V |
Input common mode range |
±5 V on ±10 mV to ±500 mV ranges
±50 V on ±1 V to ±50 V ranges |
±125 V on ±2.5 V to ±12.5 V ranges
±1000 V on ±25 V to ±1000 V ranges |
DC accuracy (10 kHz) |
±1% of full scale ±500 µV |
±3% of full scale ±12.5 mV |
Analog offset range |
±250 mV on ±10 mV to ±500 mV ranges
±2.5 V on ±1 V to ±5 V ranges
±25 V on ±10 V to ±50 V ranges |
±6.25 V on ±2.5 V to ±12.5 V ranges
±62.5 V on ±25 V to ±125 V ranges
±625 V on ±250 V to ±1000 V ranges
|
Analog offset accuracy |
1% of offset setting in addition to basic DC accuracy |
1% of offset setting in addition to basic DC accuracy |
Overvoltage protection |
±100 V DC + AC peak (any differential input to ground)
±100 V DC + AC peak (between differential inputs) |
CAT III 1000V
|