| Oscilloscope — vertical |
| Input channels |
2 |
Bandwidth
±20 mV range
±10 mV range |
5 MHz
4 MHz
3 MHz |
| Hardware bandwidth limiter |
200 kHz, switchable |
Rise time (calculated)
±20 mV range
±10 mV range |
70 ns
88 ns
117 ns |
| Input type |
Single-ended, BNC connector |
| Vertical resolution |
16 bits |
| Enhanced vertical resolution |
20 bits |
| Input sensitivity |
2 mV/div to 4 V/div (10 vertical divisions) |
| Input ranges (full scale) |
±10 mV to ±20 V in 11 ranges |
| Input coupling |
Software-selectable AC/DC |
| Input characteristics |
1 MΩ ±2% || 15 pF ±2 pF |
DC accuracy (% of full scale)
±50 mV range
±20 mV range
±10 mV range |
±0.25%
±0.5%
±1%
±2% |
| Overvoltage protection |
±50 V (DC + AC peak) |
| Oscilloscope — horizontal |
| Maximum sampling rate (real-time) |
10 MS/s |
Maximum sampling rate
(continuous streaming mode)
Using PicoScope
Using the SDK |
10 MS/s
10 MS/s one channel (PC-dependent)
6.7 MS/s two channels (PC-dependent) |
| Timebase ranges |
1 µs/div to 5000 s/div |
| Buffer memory |
16 MS shared between active channels |
| Streaming buffer memory (PicoScope) |
100 MS (shared) |
| Streaming buffer memory (SDK) |
Up to available PC memory |
| Maximum buffer segments (PicoScope) |
10,000 |
| Maximum buffer segments (SDK) |
32,768 |
| Maximum waveforms per second |
9,100 (typical) |
| Timebase accuracy |
±50 ppm |
| Sample jitter |
< 10 ps RMS |
| Dynamic performance (typical) |
| Crosstalk |
> 50,000:1 |
| Harmonic distortion |
–95 dB typical @ 10 kHz, –1 dBfs input |
| SFDR |
96 dB typical @ 10 kHz, –1 dBfs input |
| Noise |
8.5 µV RMS (on most sensitive range) |
| Bandwidth flatness |
[+0.25 dB, –3 dB] from DC to full bandwidth |
| Pulse response |
< 1% overshoot on all ranges |
|