Environmental
|
Operating temperature |
-10 °C to +50 °C |
Storage temperature |
-20 °C to +60 °C |
Power requirements |
12 V dc (Universal ac adapter/charger supplied) |
Battery |
Internal 3800 mAh advanced NiMH pack |
Operating time
|
On full charge |
Approx 50 hours (Measure only or external air; no
pump) |
Using internal pump |
Approx 100 calibration cycles to 300 psi |
Physical
|
Dimensions |
15.25 in L x 12 in W x 7 in D |
Weight |
~7 kg (15 lb) |
EMI/RFI conformance |
EN61326:2006 Annex A |
Connectors/ports |
1/8 in NPT (External supply port and test port) |
Included accessories |
Manual, NIST-traceable certificate, test leads, universal
ac adapter/charger |
Ranges
|
Pressure (internal pump) |
-80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar) |
Pressure (external air) |
0 to 2 MPa (0 to 300 psi, 0 to 20 bar) |
mA |
0 to 24.000 mA |
Volts |
0 to 30.000 V dc |
Engineering units |
psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4 °C,
mmH2O @ 20 °C, cmH2O @ 4 °C, cmH2O @ 20 °C,
inH2O @ 4 °C, inH2O @ 20 °C, inH2O @ 60 °F, mmHg
@ 0 °C, inHg @ 0 °C |
Instrumental measurement uncertainty
|
Pressure |
±(0.025 % of reading + 0.01 % FS) |
mA |
±(0.015 % of reading + 0.002 mA) |
Volts |
±(0.015 % of reading + 0.002 V) |
Temperature effect (all functions)
|
No effect on accuracy on all functions from 15 °C to 35 °C |
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C |